Easson ES-9Q Operation Manual page 42

Digital readout system

Advertisement

record down the measured value from the readout and fill up following table
standard position
10.000
20.000
30.000
40.000
50.000
.....
.....
180.000
190.000
200.000
CP.START=-90.565
CP.PITCH=10
CP.STEP=20
4) Enter the error curve data into the ES-9Q readout
switch off the readout and
then switch it on again to enter
into the "SETUP" function
11111111
6 SETUP
6 SETUP
Polar
Store
R/D
Recall
XY
11111111
11111111
Q
Non-linear error compensation
during the start up of readout,
when the readout displays
the software version number,
press "6" once to enter
int the "SETUP" function
Print
Enter
FI N I S H
measured result
10.003
20.007
30.005
40.010
50.007
.....
.....
180.009
190.090
200.050
Q
- 36 -
SETUP
Polar
Store
R/D
Recall
Print
XY
Enter
FI N I S H