record down the measured value from the readout and fill up following table
standard position
10.000
20.000
30.000
40.000
50.000
.....
.....
180.000
190.000
200.000
CP.START=-90.565
CP.PITCH=10
CP.STEP=20
4) Enter the error curve data into the ES-9Q readout
switch off the readout and
then switch it on again to enter
into the "SETUP" function
11111111
6 SETUP
6 SETUP
Polar
Store
R/D
Recall
XY
11111111
11111111
Q
Non-linear error compensation
during the start up of readout,
when the readout displays
the software version number,
press "6" once to enter
int the "SETUP" function
Print
Enter
FI N I S H
measured result
10.003
20.007
30.005
40.010
50.007
.....
.....
180.009
190.090
200.050
Q