Troubleshooting
October 2021
8.3
NQ Ratio above the permitted limit
Host status = F
0-20 = 1 mA
Light-emitting diode (LED) = yellow, blinking
Table 8-3: Potential Cause and Resolution
Potential cause
Poor alignment
8.4
Ratio out of the limit
Host status = F
0-20 = 1 mA
Light-emitting diode (LED) = yellow, blinking
Table 8-4: Potential Causes and Resolutions
Potential cause
Poor alignment
Dirt on the window
Detector fault
8.5
Voltage less than 16 Vdc
The detector is at V fault.
Table 8-5: Potential Cause and Resolution
Potential cause
Low input voltage
62
Resolution
Align.
Resolution
Align.
Clean the window.
Replace or repair the detector.
Resolution
Check the power supply and installation.
Emerson.com/Rosemount