The battery pack has an estimated life of 300–500 charge-discharge cycles,
depending on use and storage conditions.
Battery life can be extended by storing in cool conditions. High temperatures
can cause accelerated aging.
To recondition the battery, put it on the charging dock for 12 hours. The dock will
fast charge the battery and the additional trickle charge time will recondition it
for maximum capacity. Remove the pump from the dock and let it rest for at
least one hour. Place it on the dock for another charge cycle.
As the battery goes through charge/discharge cycles and ages, its capacity will
diminish. If the battery is reconditioned as described above and does not result
in expected run times, the battery has reached end of life and should
be replaced.
7.2

Battery replacement

The battery replacement process is as follows:
You will require a replacement battery (part no. SERFFP-1004), and a
NOTES
cross-head screwdriver.
[1]
Remove the three case screws
Remove the battery pack from the front
[2]
The battery pack is connected to the PC board by a wire harness.
NOTES
Carefully unplug the wire harness, noting the position of the six pin connector.
[3]
[4]
Connect the new battery, ensuring the correct pin pattern.
Reconnect the case halves.
[5]
Pay attention to wire routing to ensure not to pinch the wire harness.
Replace the case screws, and snug tighten only. Tighten until the gap between
[6]
the case halves fully closes.
Step 1
Page 20
Step 2
Step 3-4
QUI-1164
Step 5-6