Control Panel; Function Keys - Cantel Medical MEDIVATORS CER OPTIMA User Manual

Endoscope reprocessing system
Table of Contents

Advertisement

Control Panel

The control panel allows the operator to select a cycle, adjust cycle settings, view system and cycle status, view
alarm conditions, and fully operate the reprocessor.

Function Keys

The function keys control the operation of the CER OPTIMA
AUTOMATIC Cycles
In the AUTOMATIC mode, the CER OPTIMA system will control all reprocessing cycle functions. Two
different AUTOMATIC functions are available depending on the type of cycle the operator wants to perform.
FULL Cycle: Sets the reprocessing cycle to automatically provide a detergent wash/flush, rinse, HLD
exposure, post-rinses (HLD dependent) and air and alcohol endoscope flushing (if selected).
The HLD exposure time is pre-set based on the HLD in-use which also determines the number of post
rinses required (per the HLD manufacture's recommendations).
The WASH time is pre-selected by the operator:
The CER-1 OPTIMA allows a 2, 3 or 5 minutes wash time
The CER-2 OPTIMA allows a 3 or 5 minute wash time
If a detergent wash is desired, then the detergent is manually dispensed by the operator during the
WASH phase of the cycle.
The AIR time and ALCOHOL volume is pre-selected by the operator when setting up the cycle.
The AIR time is selectable from 0, 3, 5, and 10 minutes
34
Figure 10. Control Panel
endoscope reprocessor.
®