SimMan 3G PLUS Important Information
temperature (moving the Patient Simulator
from a cold environment to a warm
environment and vice versa) may result in
condensation collecting on the base board
and pose a shock hazard.
• Defibrillation must be performed on the
defibrillator discs only.
• The Patient Simulator must not come into
contact with electrically conductive surfaces
or objects during defibrillation.
• Do not defibrillate the Patient Simulator in a
flammable or oxygen enriched atmosphere.
• The ECG connectors are designed exclusively
for ECG monitoring and must not be
used for defibrillation. Defibrillation on the
ECG connectors will damage the internal
electronics of the Patient Simulator and may
cause personal injury.
Cautions
• Using a defibrillator in temperatures over
35 °C (95 °F) may cause overheating and
shutdown. If automatic shutdown occurs, allow
the Patient Simulator to cool down before
resuming the training session. Open the torso
skin to speed up the cooling process.
• To prevent torso skin electrode pitting, do
not apply conductive gel or conductive
defibrillation pads intended for patient use.
Battery
Warnings
• Inserting and connecting batteries incorrectly,
short circuiting or exposure to fluids pose an
explosion hazard.
• Do not mistreat, disassemble or attempt to
repair the battery.
• Do not use the batteries if they are visibly
damaged, malfunction or appear to leak
electrolyte.
• Take extreme care to avoid direct contact
with electrolyte, hot or smoking parts. In case
of the above, disconnect and remove the
battery when it is judged safe to do so.
• Only use the SimMan Family (except
SimMan ALS) external power supply and
batteries.
• The external battery charger is for indoor
use only.
• Never store fully charged batteries for longer
than a month.
• Never store the batteries inside the Patient
Simulator.
• The batteries should only be charged and
used at the specified temperature ranges.
Cautions
• Do not run the Patient Simulator for more
than 1 minute on a single battery.
• After the Patient Simulator is turned off, wait
20 seconds before restarting or the Patient
Simulator may not function properly.
For other battery related information, consult
User Guide.
Symbol Glossary
CE mark
Australia Radiocommunications
and EMC Compliance Mark
Waste Electrical and Electronic
Equipment symbol
Manufacturer
Date of Manufacture
Reference Number
Serial Number
Li-Ion batteries recycling symbol
Warranty
Refer to the Laerdal Global Warranty for
terms and conditions. For more information
visit www.laerdal.com
4