Safety Information

Gas-Pro is a hazardous area certified gas detector and as such must be operated and
maintained in strict accordance with the instructions, warnings and label information included in
this manual. Gas-Pro must be operated within the limitations stated.
Read and understand all instructions in the operation section of this manual prior to use.
Before use ensure that the equipment is in good condition, the enclosure is intact has not been
damaged in any way.
If there is any damage to the equipment do not use, contact your local office or agent for repair/
replacement.
Do not disassemble or substitute components as this may impair intrinsic safety and invalidate
safety certification.
Only genuine replacement parts must be used; substitute components may invalidate
certification and warranty of the Gas-Pro and accessories, reference "Service and
Maintenance" section for details.
No live maintenance is permissible.
Observe all warnings and instructions marked on the unit and within this manual.
Observe site health and safety procedures for gases being monitored and evacuation procedures.
Understand the screen display and alarm warnings prior to use.
If this product is not working properly, read the troubleshooting guide and/or contact your local
office or agent, for details reference the 'Contacts' section of the manual (see
Ensure maintenance, service and calibration is carried out in accordance with the procedures in
the manual and only by trained personnel.
The Gas-Pro re-chargeable battery must only be charged in non-hazardous (safe) areas.
Only connect to Gas-Pro in a safe area for charging or communications.
Gas-Pro must not be charged or have communication to the device, at ambient temperatures
greater than +40°C.
Charging cable assemblies, whilst nominally providing a voltage of 6.5V, must not exceed a
voltage of 9.1V as this may impair intrinsic safety and invalidate safety certification (Um=9.1V).
Communication cable assemblies, whilst nominally operating at a voltage of 3.0V TTL, must not
exceed a voltage of 9.1V as this may impair intrinsic safety and invalidate safety certification
(Um=9.1V).
The devices are intended for use in normal atmospheric conditions of temperature –20 °C to
+55 °C; pressure 80 kPa (0,8 bar) to 110 kPa (1,1 bar); and air with normal oxygen content,
typically 21 % v/v (volume/volume).