OBLIGATIONS OR LIABILITIES ON THE PART OF RKI INSTRUMENTS, INC. INCLUDING BUT NOT LIMITED TO, THE WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. IN NO EVENT SHALL RKI INSTRUMENTS, INC. BE LIABLE FOR INDIRECT, INCIDENTAL, OR CONSEQUENTIAL LOSS OR DAMAGE OF ANY KIND CONNECTED WITH THE USE OF ITS PRODUCTS OR FAILURE OF ITS PRODUCTS TO FUNCTION OR OPERATE PROPERLY.
Page 5
WARNING: Do not use this product in a manner not specified in this instruction manual. WARNING: When using the 65-2330RK, you must follow the instructions and warnings in this manual to assure proper and safe operation of the 65-2330RK and to minimize the risk of personal injury.
Page 6
Controller Terminal Strip Factory Adjust Pot Span Pot Zero Pot Amplifier Detector Terminal Strip + Test Point (Red) 100 mV - 500 mV Range - Test Point (Black) J-Box H2S Detector GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 7
Detector Housing Body Cap Gasket Plug-In H2S Sensor Detector Housing Cap Hydrophobic Membrane Detector Housing Body Housing Cap & Cap Gasket Plug-In H S Sensor GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 8
Controller Terminal Strip Detector Terminal Strip Zero Pot Span Pot CAUTION:The third potentiometer is factory-set. Do not adjust it. Test Points GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 9
3.10 3/4 NPT Female 3.65 6.8 Max Rubber Spacers J-Box H2S Detector 1 1 /2-20 T hread for Calibration Cup GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 10
WARNING: Always verify that power to the controller is OFF before you make wiring connections. CAUTION:If shielded cable is used, leave the cable shield’s drain wire insulated and disconnected at the transmitter. You will connect the opposite end of the cable’s drain wire to the controller’s chassis (earth) ground.
Page 11
4 - 20 mA In (S) + 24 VDC Cable Shield Controller J-Box Black H2S Detector GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 12
CAUTION:Allow the transmitter to warm up for 5 minutes before you continue with the next section, “Setting the Zero Signal”. GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 14
Note: Under “normal” circumstances, the transmitter requires calibration once every 3 months. Some applications may require a more frequent calibration schedule. Replacing the Plug-In H S Sensor CAUTION:The sensor contains electrolyte which is a dilute acid. Do not disassemble the sensor when replacing it with a new one.
Page 15
WARNING: The H S sensor will not operate properly if the wire jumper is not removed. Remove jumper CAUTION:Allow the replacement sensor to warm up for 5 minutes before you continue with the next step. Replacing the H S Detector GlobalTestSupply www.
Page 16
CAUTION: Allow the replacement detector to warm up for 5 minutes before you continue with the next step. Replacing the Hydrophobic Membrane Replacing the Amplifier GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 18
WARNING: Do not remove the detector housing cap or junction box cover while the circuits are energized unless the area is determined to be non-hazardous. Keep the detector housing cap and junction box cover tightly closed during operation. Output (mV) = (calibrating sample/fullscale) X 400 + 100 200 (mV) = (25/100) X 400 +100 GlobalTestSupply www.