RKI Instruments 65-2320RK-TFX-4 Operator's Manual

Oxygen transmitter

Advertisement

Quick Links

65-2320RK-TFX-4
Oxygen Transmitter
Operator's Manual
Part Number: 71-0669
Revision: P1
Released: 4/2/24
RKI Instruments, Inc.
www.rkiinstruments.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 65-2320RK-TFX-4 and is the answer not in the manual?

Questions and answers

Summary of Contents for RKI Instruments 65-2320RK-TFX-4

  • Page 1 65-2320RK-TFX-4 Oxygen Transmitter Operator’s Manual Part Number: 71-0669 Revision: P1 Released: 4/2/24 RKI Instruments, Inc. www.rkiinstruments.com...
  • Page 2 For most applications, typical calibration frequencies are between 3 and 6 months but can be more often or less often based on your usage. 2 • 65-2320RK-TFX-4 Oxygen Transmitter Operator’s Manual...
  • Page 3 We do not assume indemnification for any accident or damage caused by the operation of this gas monitor and our warranty is limited to the replacement of parts or our complete goods. Warranty covers parts and labor performed at RKI Instruments, Inc. only, and does not cover field labor or shipment of parts back to RKI.
  • Page 4: Table Of Contents

    Parts List ............23 4 • 65-2320RK-TFX-4 Oxygen Transmitter Operator’s Manual...
  • Page 5: Overview

    Accuracy ± 0.5% O WARNING: When using the 65-2320RK-TFX-4, you must follow the instructions and warnings in this manual to assure proper and safe operation of the 65-2320RK- TFX-4 and to minimize the risk of personal injury. Be sure to maintain and periodically calibrate the 65-2320RK-TFX-4 as described in this manual.
  • Page 6: Description

    3/4 NPT conduit fitting. Two color-coded leads extend from the top of the detector. The leads allow you to connect the detector to a controller. 6 • Description 65-2320RK-TFX-4 Oxygen Transmitter Operator’s Manual...
  • Page 7: Calibration Adapter/Splash Guard

    Detector Terminal Strip The detector terminal strip is a four-point plug-in style terminal strip near the bottom of the amplifier. Use the detector terminal strip to connect the oxygen detector to the amplifier. 65-2320RK-TFX-4 Oxygen Transmitter Operator’s Manual Description • 7...
  • Page 8: Amplifier Junction Box

    A terminal block within the junction box facilitates the wiring process. A cover on the front of the junction box allows access to the interior of the junction box. 8 • Description 65-2320RK-TFX-4 Oxygen Transmitter Operator’s Manual...
  • Page 9: Installation

    2. Amplifier junction box dimensions are shown in Figure 3. Detector junction box dimensions are shown in Figure 4. Ø3.65 2.70 3/4 Conduit Hub 5.25 Rubber Spacer, 3/4 Conduit Hub Figure 3: Amplifier Junction Box Dimensions 65-2320RK-TFX-4 Oxygen Transmitter Operator’s Manual Installation • 9...
  • Page 10 4. Install 3/16 inch I.D. flexible polyurethane tubing to the fitting at the bottom of the calibration adapter/splash guard and route it to an accessible area that is close to the amplifier junction box. See the “Parts List” on page 23 for available tubing. 10 • Installation 65-2320RK-TFX-4 Oxygen Transmitter Operator’s Manual...
  • Page 11: Wiring The Detector To The Amplifier

    To make wiring more convenient, use wire colors that correspond to the detector wire colors: white and green. NOTE: Consult RKI Instruments, Inc. for cable lengths longer than 200 feet. 5. Connect the wires to the terminal block in the detector junction box.
  • Page 12 Detector J-Box Oxygen Detector Calibration Adapter/Splash Guard Figure 5: Wiring Detector to Amplifier 10. Reinstall the detector terminal strip into its socket. 11. Secure the amplifier junction box’s cover to the junction box. 12 • Installation 65-2320RK-TFX-4 Oxygen Transmitter Operator’s Manual...
  • Page 13: Wiring The Amplifier To A Controller

    CAUTION: Do not route power and transmitter wiring through the same controller conduit hub. The power cable may disrupt the transmission of the transmitter signal to the controller. 65-2320RK-TFX-4 Oxygen Transmitter Operator’s Manual Installation • 13...
  • Page 14: Startup

    2. Verify that the power wiring to the controller is correct and secure. Refer to the controller operator’s manual. 3. Turn on power to the controller. 4. Turn on the controller. 14 • Start Up 65-2320RK-TFX-4 Oxygen Transmitter Operator’s Manual...
  • Page 15: Setting The Fresh Air Signal

    9. If necessary, use a small flat-blade screwdriver to adjust the span pot until the voltmeter reading is 434 mV (±2 mV). 10. Turn the regulator’s on/off knob clockwise to close it. 65-2320RK-TFX-4 Oxygen Transmitter Operator’s Manual Start Up • 15...
  • Page 16: Maintenance

    The troubleshooting guide describes symptoms, probable causes, and recommended action for problems you may encounter with the oxygen transmitter. NOTE: This troubleshooting guide describes transmitter problems only. See the controller operator’s manual for problems you may encounter with the controller. 16 • Maintenance 65-2320RK-TFX-4 Oxygen Transmitter Operator’s Manual...
  • Page 17: Replacing Components Of The Oxygen Transmitter

    Replacing the Oxygen Detector 1. Turn off the controller. 2. Turn off or unplug power to the controller. 3. Unscrew the calibration adapter/splash guard from the detector. 65-2320RK-TFX-4 Oxygen Transmitter Operator’s Manual Maintenance • 17...
  • Page 18 7. Place the new amplifier in the same position as the old amplifier. A foam gasket that orients the amplifier and keeps it from rotating is installed on the bottom of the amplifier. Make sure the amplifier is seated flat in the junction box. 18 • Maintenance 65-2320RK-TFX-4 Oxygen Transmitter Operator’s Manual...
  • Page 19: Calibration Frequency

    Although there is no particular calibration frequency that is correct for all applications, a calibration frequency of every 3 months is adequate for most oxygen transmitter applications. Unless experience in a particular application dictates otherwise, RKI Instruments, Inc. recommends a calibration frequency of every 3 months for the oxygen transmitter.
  • Page 20: Calibration

    It describes the test using a calibration kit that includes calibration gas and a fixed flow regulator with an on/off knob. RKI Instruments, Inc. recommends using a 0.5 LPM (liters per minute) fixed flow regulator.
  • Page 21: Setting The Fresh Air Reading

    4. Verify a reading of 100 mV (± 2mV). If necessary, use the zero pot on the amplifier to adjust the reading to match the correct response reading. 5. Turn the regulator knob clockwise to close the regulator. 6. Unscrew the regulator from the calibration cylinder. 65-2320RK-TFX-4 Oxygen Transmitter Operator’s Manual Calibration • 21...
  • Page 22: Returning To Normal Operation

    4. When the display reading rises above the decreasing alarm’s setpoint, return the controller to normal operation. 5. Verify that the controller display reading decreases and stabilizes at 20.9% volume. 6. Store the components of the calibration kit in a safe and convenient place. 22 • Calibration 65-2320RK-TFX-4 Oxygen Transmitter Operator’s Manual...
  • Page 23: Parts List

    S2 series oxygen amplifier with orienting gasket 65-2510RK-TFX Oxygen replacement detector assembly (not including detector junction box), CSA classified 71-0669 65-2320RK-TFX-4 Oxygen Transmitter Operator’s Manual (this document) 81-0076RK-01 Zero air calibration cylinder, 34 liter steel 81-0078RK-01 100% nitrogen calibration cylinder, 34 liter steel...