3.4 Changing the Air Filter

Change the air filter every 60 days even if it does not appear dirty. If it appears dirty, or water or
medication has spilled on it, replace with a new one immediately.
1
Pull the air filter cover off the main unit.
2
Remove the dirty air filter.
Note
Do not wash or clean the air filter. The dirty air filter should be disposed of.
3
Insert a new air filter.
CAUTION
• Use ONLY Omron authorized parts and accessories (Refer to section 6). Parts and accessories
not approved for use with this device may damage the unit.
• Make sure the air filter is clean. If the air filter has changed color or has been used on average
for more than 60 days, replace it with a new one.
Note
• Do not substitute cotton or any other material for the filter.
• Before inserting the new air filter check to make sure it is clean and free of dust. Do not operate
the device without the air filter.
4
Put the air filter cover back in place. Make sure it is attached properly.
17