R ou t in e S af e ty T es ti n g an d In s pe ct io n

The follo wing electr ical tests and in spection info rm ati on is provide d fo r gu idance purpo ses only.
These tests involve the use of volta ges and cu rr ents that can cause inj ury a nd shou ld onl y be
per formed by qu alifie d per sonnel familia r w i th ES D and e lectr ical saf ety p recaut ions.
Pr ior to ca rryi ng out any inspecti on or te st pr oc edur e, d isconnect all e xtern al eq uipment fro m the
Test Set and disconne ct th e Test Se t f rom t he A C Pow er Supp ly. All tests should include the
instr ument' s ow n supply lead, all covers mu st be fit ted a nd t he 390 0 A C P ow er Suppl y Sw itch must
be i n the ON positio n.
Reco mm e nded t ests a nd insp ection should be car ried out in t he f ollow ing sequence:
Visua l In sp ection
Ear th (Grou nd) Bondi ng Te sts
Insula tion Resist ance T est
E a r th B o n d i n g T es ts
Ear th bonding test s should be p erf ormed using a 25 A ( 12 V maximum o pen cir c uit voltag e) D C
source . Tests sh ould b e limit ed t o a ma xi mu m dur ation of 5 seconds and h ave a pass limit of 0.1
ohm aft er allow ing fo r the r esistance of the supply lea d. Exceedin g the 5 second te st durat ion can
damage t he Test Set . Th e tests sh ould be ca rri ed out betw een t he supply ea rth and expose d case
metalw or k. N o attempt shoul d be made t o per for m the test s on functi onal e arth s (e. g., signal
carr ying co nnector shell s or screen connec t ions) as t his damag es the Test Set .
I n su l a ti o n T e st s
A 500 V DC te st sho uld b e app lied betw een the prot ective ear th co nnectio n and combine d live and
neut ral suppl y conn ections w hile the 3 900 AC Pow er S upply Sw itch is in th e ON position . T o avoid
swit ching the live and neut ral poles on th e Test S et, establi sh the live/ neutr al l inks on the
applia nce te ster or its co nnector . The test voltag e should be applied for 5 se conds be fore takin g
the measure me nt.
Aer ofle x uses r einf orced insulat ion in the constr uction of the 3900 and, ther efor e a minimum pass
limit of 7 Mohm sho uld be achieve d dur ing t his t est.
When a D C pow er a dapter is provid ed w ith the equipment , t he a dapter must pass t he 7 Mohm te st
limit.
Aer ofle x does not r ecommend dielect ric f lash t esting dur ing r outi ne saf ety t ests. Most port able
applia nce te sters use A C for the dielect ric stren gth test which can d amage t he supp ly inpu t f ilter
capacit ors.
R ec o rd Ma in te n an c e
Aer ofle x recommen ds tha t t est r esults be r ecord ed and revi ewed as pa rt of e ach te st insp ection.
Sig nificant dif fer ences be twe en pr evious readi ngs and measur ed valu es should be investiga ted.
If any f ailur e is d etecte d dur ing vi sual in specti on or electr ical tests, the equipme nt sho uld be
disable d and evaluat ed by a qu alified servi ce tech nician.
Saf ety cr itical componen ts shou ld onl y be r eplace d w ith co mparab le par ts, using techniq ues and
pro ce dures recommend ed by Aer oflex.
The above i nfor ma tion is pr ovided f or guidance pur pos es onl y. A ero flex d esigns an d manuf actur es
its p rodu ct s in accorda nce w ith Inte rnat ional Safe ty S tanda rds so that w hen u sed in accorda nce
wit h r ecommend g uidelin es, t hey r eprese nt n o hazar d t o th e oper ator . Aer ofle x rese rves t he r ight
to amend t he ab ove inf ormat ion i n the cour se of its co ntinui ng commit me nt t o pr oduct safet y .
In stal lat ion
S u b j e ct t o E x p o r t C o n tr o l , s ee C o ve r P a g e f o r d e t ai l s .
2- 12