Performance Tests
Performance Test Instructions
4 Press A
. You should hear a "beep" and the Configuration
UX
menu should disappear.
5 Press A
. You should hear a "beep".
PPL
6 Press H
. You should hear a "beep".
ELP
7 Press P
, the green hardkey. An egg-timer icon should be
RESET
displayed while all parameters are set to their default values. The
Overview Screen showing all channels should appear.
8 Press C
several times.
HANNEL
The highlighted field on the display should toggle between the
channels you have inserted a module into.

Testing the Softkeys and the Cursor Key

9 Press the right arrow of the cursor key several times.
Check the fields on the display being highlighted one after
another.
10 Press the left arrow on the cursor key several times.
Check the fields on the display being highlighted one after
another.
11 Press the up arrow on the cursor key several times.
Check the fields on the display being highlighted one after
another.
12 Press the down arrow on the cursor key several times.
Check the fields on the display being highlighted one after
another
13 Using the cursor key, move to the power parameter, [P], of the
Power Sensor module.
Testing the Softkeys and the E
14 Press [Menu]. The menu should appear.
15 Press [Close]. The menu should disappear.
16 Press [Details]. The details screen should appear. You should see
the [Cal], [Ref], and [Range] parameters in addition to [P] and
[λ]. If you use the HP 8164A Lightwave Measurement System,