C.1 HP 8163A Specifications
Display
Compatibility with HP 8153A
Lightwave Multimeter Modules
Data Acquisitions
Selectable data averaging time
Selectable total data averaging time
Trigger
Environmental
Storage temperature
Operating temperature
Power
Dimensions
Weight
HP-IB interface
HP-IB capability
Appendix C. Specifications
HP 8163A Specifications
The HP 8163A Lightwave Multimeter has a dual channel display.
Graphical display 190 × 300 points visible,
monochrome.
The HP 8163A mainframe supports all modules
from the HP 8153A Lightwave Multimeter, apart
from the HP 81534A Return Loss Module .
Memory>
4000 measurement results / channel
min 100 µs or 10 ms, depending on sensor module
Sampling time
20 ms to 60 min
20 ms to 99:59:59 h
True synchronous on all channels
−40°C to +70°C
0°C to +45°C
Humidity
<95% R.H. from 0°C to +45°C
AC 100 - 240 V
88 mm × 213 mm × 380 mm
(3.5" × 8.4" × 15")
net 4.200 kg (9.3 lbs); shipping 6.5 kg (14.5 lbs)
Function code
SH1, AH1, T6, L4, SR1, RL1, PP0, DC2, DT0, C0.
SCPI standard; HP 8153A code compatibility
All modes and parameters accessible via HP-IB
interface
169
±
10%, 48 - 66 kHz, 100 VA max.