Status Reporting – The STATus Subsystem
command:
syntax:
description:
parameters:
response:
example:
command:
syntax:
description:
parameters:
response:
example:
command:
syntax:
description:
parameters:
response:
example:
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User's Guide, E1299
:STATus:OPERation:ENABle:LEVel1?
:STATus:OPERation:ENABle:LEVel1?
Returns the OSESM for the OSESR for slots 15 - 17 of the HP 8166A Lightwave Measure-
ment System
none
The bit value for the operation enable mask as a 16-bit signed integer value (0 .. +32767)
stat:oper:enab:level1? → +128<END>
:STATusn:OPERation[:EVENt]?
:STATusn:OPERation[:EVENt]?
Returns the Operational Slot Status Event Register (OSSER) of slot n.
none
The results for the individual slot events (a 16-bit signed integer value, where 0 ≤ value ≤
32767):
Bit
Mnemonic
4-15
Not used
3
Slot n: Zeroing started
2
Not used
1
Slot n: Coherence Control has been switched on
0
Slot n: Laser has been switched on
stat1:oper? → +0<END>
:STATusn:OPERation:CONDition?
:STATusn:OPERation:CONDition?
Returns the Operational Slot Status Condition Register of slot n.
none
The results for the individual slot events (a 16-bit signed integer value, where 0 ≤ value ≤
32767):
Bit
Mnemonic
4-15
Not used
3
Slot n: Zeroing ongoing
2
Not used
1
Slot n: Coherence Control is switched on
0
Slot n: Laser is switched on
stat1:oper:cond? → +0<END>
Instrument Setup and Status
Decimal Value
0
8
0
2
1
Decimal Value
0
8
0
2
1