Before installing the lightwave switch, refer to "General Safety
Considerations" on page -iii of this manual.
OPTION 3XX INSTRUMENTS: To avoid damage, handle the pigtail fiber with
care. Use only an appropriate fiber cleaver tool for cutting the fiber. Do not pull
the bare fiber out of its jacket, crush it, kink it, or bend it past its minimum bend
radius.
Install the instrument so that the front panel ON/OFF switch is readily identi-
fiable and is easily reached by the operator. The ON/OFF switch, or the
detachable power cord, is the instrument disconnecting device. It disconnects
the mains circuit from the mains supply after the EMC filters and before other
parts of the instrument. Alternatively, an external installed switch or circuit
breaker (which is easily identifiable and is easily reached by the operator)
may be used as a disconnecting device.
Install the instrument according to the enclosure protection provided. This
instrument does not protect against the ingress of water. This instrument does
protect against finger access to hazardous parts within the enclosure.