Page 4 particles. ADJUSTMENTS Page 5 THE DEVICE Page 7 The CDM controls in real time the dust emission rate and DYSFUNCTION Page 9 warns if this rate becomes higher than the ALARM and CHARACTERISTICS Page 10 ALERT level, adjusted by the operator on site.
CDM-Sensor – Dust Emission Incident Detector MECHANICAL ASSEMBLY The sensor is directly fixed on the pipe to monitor, in the following conditions: Thread G1/2" -placed on the straight part of the pipe, following the drawing underneath, -on a metallic pipe linked to earth, we must have a continuity of earth between the sensor and the pipe, -standard length of the probe: 220 mm for 250mm <...
Warning: Connect the sensor to earth by the terminal which is on the body of the sensor. Check the electrical characteristics of the CDM before any connection to the process, to prevent any damage to the CDM. We recommend shielded cables between the sensor and the COM RS485.
Adjustment of the Value X (from 1 to 33 min): Value in minutes. During this time the CDM defines a mean value of the dust rate, in MEASURE mode and AUTOLEARNING mode of the PR. This mean value is recalculated each second.
Page 6
*: Value defined in our workshop. Remarks: In case of supply shutdown, the CDM doesn’t need to be adjusted again. All the adjusted parameters are recorded. The adjustments, which need the opening of the cover, must be done quickly. Don’t let the cover open when not needed.
CDM-Sensor – Dust Emission Incident Detector THE DEVICE User Interface Figure 4 : user interface Supply Led ALERT Led ALARM Led MES/n2 Led 11 parts BARGRAPH (0 to 100%) TST/n1 Led PR/x LED Key F Key + SEL1, SEL2 Notice Technique FI 72.0401.0819E...
FAILURE The CDM regularly runs an AUTOTEST of its electronics, to check its efficiency. If an internal failure is detected by the CDM, this one informs the user by the ALARM Led and the ALARM Relay. Table 3 : ALARME LED...
If Ø < 250mm : on request If Ø > 1000mm : consult us Running Temperature - external part to the pipe, body of the CDM : -20°C to +60°C - internal part to the pipe : flow < 150° (CDM standard) Storage Temperature -20°C to +85°C...
In the event of a problem, please contact Sefram. II. RECEPTION - STORAGE V. MAINTENANCE After unpacking the device, check that this latter has not The device does not require any special maintenance.
PLACE GUTENBERG 59175 TEMPLEMARS CERTIFICATE The CDM respects the European directives ( CEM and ATEX ), which concerns it. However, it must be used correctly in applications for which it is intended, and should be linked or near CE approved products.
Need help?
Do you have a question about the CDM and is the answer not in the manual?
Questions and answers