Siargo AM1000 Series User Manual

Siargo AM1000 Series User Manual

Insertion mems flow, temperature and humidity sensors

Advertisement

Quick Links

AM1000 User Manual
VB.0.1
Insertion MEMS flow, temperature and humidity sensors
©2021 Siargo Ltd.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the AM1000 Series and is the answer not in the manual?

Questions and answers

Summary of Contents for Siargo AM1000 Series

  • Page 1 AM1000 User Manual VB.0.1 Insertion MEMS flow, temperature and humidity sensors ©2021 Siargo Ltd.
  • Page 2 Copyright 2021 and Liability Disclaimer Siargo Ltd. and its subsidiaries reserve the right to change the specifications and/or descriptions without prior notice. Siargo and its subsidiaries shall not assume any inaccuracy or errors in this manual. For further information and updates, please visit www.Siargo.com.
  • Page 3 • Do not apply for any unknown or non-specified gases that may damage the product. www.Siargo.com AM1000 User Manual...
  • Page 4: Table Of Contents

    8.1. Measurement principles ........................19 8.2. Precautions for the best performance of the product ............... 19 8.2.1. Altitude changes ..........................19 8.2.2. Excessive humidity or condensation ..................... 19 8.2.3. Metrology verification ........................20 9. Warranty and Liability ..........................21 www.Siargo.com AM1000 User Manual...
  • Page 5 10. Service contact and information ........................ 23 Appendix I: Sensor evaluation kit ........................24 Appendix II: Document history ........................25 www.Siargo.com AM1000 User Manual...
  • Page 6: Overview

    All contact information can be found at the end of this manual. This manual provides essential information for the AM1000 series of integrated MEMS sensors of mass flow (flow speed), temperature, and humidity. It can be readily applied for environmental monitoring as well as an insertion flow meter.
  • Page 7: Receipt / Unpack Of The Products

    Please note that the sensor has the standard 1.0 meter data and power cable directly attached to the product, therefore no separate cable is needed for the operation. Please follow the instructions in the subsequent section for the correct connections for the exact model you ordered. www.Siargo.com AM1000 User Manual...
  • Page 8: Knowing The Products

    C or RS85 Modbus Halfplex that can be selected at the time of order. These two communication protocols share the ports as defined in Table 3.2. For the detailed protocols of the corresponding option, please refer to Section 5. www.Siargo.com AM1000 User Manual...
  • Page 9: Mechanical Dimensions

    Φ12.5 mm. A flat plateau hard materials attached to the pipe should be prepared to fix the probe to the pipe, refer to the following graphs. The fixtures should be leakage-proof tested before the installation of the product. An illustration sketch is shown below: www.Siargo.com AM1000 User Manual...
  • Page 10 Figure 4-2. Installation with correct upstream and downstream space. L = 1/8D+H+B+23 where D is the flow channel diameter H is the flow channel wall thickness B is the fixture base height. Figure 4-3. Insertion probe length calculation. www.Siargo.com AM1000 User Manual...
  • Page 11 The product outputs the flow speed. In order to calculate the flow rate, it is necessary to know the depth that the product is inside the flow channel. Please refer to the above graph for the calculations. www.Siargo.com AM1000 User Manual...
  • Page 12: Digital Communication Descriptions

    Note: 1. The I C address is set to Bit 7~Bit 1. e.g. if the I C address is 4 (0000 100x), the write address will be 0x08 (0000 10o0) and the read address will be 0x09 (0000 1001). www.Siargo.com AM1000 User Manual...
  • Page 13: I 2 C Interface Read/Write Sequences

    5.1.3 I C interface read/write sequences www.Siargo.com AM1000 User Manual...
  • Page 14: Rs485 Modbus Communication Protocol

    38400 bps Start bits Data bits Stop bits Even/Odd parity None Bits period 104.2 µsec Bytes period 1.1458 msec Maximum data length Maximum nodes 5.2.2. Frame The frame function is based on the standard Modbus RTU framing: www.Siargo.com AM1000 User Manual...
  • Page 15: Function Codes

    Note: At the time of shipping, the write protection function is enabled except for address and baud rate. Once the user completes the register value change, the write protection will be automatically enabled once again to prevent incidental data loss. www.Siargo.com AM1000 User Manual...
  • Page 16 0x0025 Read Description Measured temperature data Value type UINT 16 Notes Temperature = value (0x0025) /100 °C Write Humidity 0x0026 Read Description Measured humidity data Value type UINT 16 Notes Humidity = value (0x0026) /100 %RH. www.Siargo.com AM1000 User Manual...
  • Page 17: Product Selection And Order Information

    For large quantities, please contact the sales office, distributors, or sales representatives. Siargo is making every effort to ensure the quality of the products. In case of questions and/or product supports, please contact the customer service listed at the end of the document.
  • Page 18: Product Performance

    C / RS485; Analog: 0.5~4.5Vdc Cable 5 color coded, 1.0m Storage temperature -20~70 °C Calibration Air @ 20°C, 101.325kPa Compliance RoHS IEC 61326-1;-2;-3 *Note: 1. Analog output only applies to flow speed. Temperature and humidity data are via digital interface. www.Siargo.com AM1000 User Manual...
  • Page 19: Technical Notes For The Product Performance

    8.2.2. Excessive humidity or condensation The humidity change will not alter the performance of the sensor. However, if excessive humidity is present resulting in condensation, the measurement channel could be blocked or altered. This would www.Siargo.com AM1000 User Manual...
  • Page 20: Metrology Verification

    For the detailed information, please contact the manufacture for the technical while paper. www.Siargo.com AM1000 User Manual...
  • Page 21: Warranty And Liability

    The products returned under warranty to Siargo shall be at the user or purchaser's risk of loss and will be returned, if at all, at Siargo's risk of loss. Purchasers or users are deemed to have accepted this limitation of warranty and liability, which contains the complete and exclusive limited warranty of Siargo.
  • Page 22 (5) Any damages incurred by the incorrect usage of the products; (6) Siargo does not provide any warranty on finished goods manufactured by others. Only the original manufacturer's warranty applies;...
  • Page 23 10. Service contact and information Siargo Ltd. is making every effort to ensure the quality of the products. In case of questions and or product supports, please contact customer service at the address listed below. We will respond to your request in a timely fashion and work with you toward your complete satisfaction.
  • Page 24 Appendix I: Sensor evaluation kit Siargo offers a sensor evaluation kit, including a digital data converter, USB data cable, and User Application software, that allows the user to evaluate the product performance on a Microsoft Windows-based computer. The user can read and visualize the flow rate of the product, obtain the totalized values, and save the data for further analysis.
  • Page 25 ➢ Revision A.8 (July 2018): C communication update: temperature and humidity. ➢ Revision A.7 (May 2018): C communication update: sequences. Sales Europe: IDENTIC GmbH In der Siedlerruh 24 69123 Heidelberg /Germany Phone: +49 6221 7509 777 Email: info@identic.de www.Siargo.com AM1000 User Manual...

Table of Contents