OXYGEN CONCENTRATORS
Operators Manual
Model OC8000 & OC8200
M8000 (not shown)
For Veterinary Use Only
Supera , LLC Copyright 2020

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Pureline OC8000 and is the answer not in the manual?

Questions and answers

Summary of Contents for Supera Pureline OC8000

  • Page 1 OXYGEN CONCENTRATORS Operators Manual Model OC8000 & OC8200 M8000 (not shown) For Veterinary Use Only Supera , LLC Copyright 2020...
  • Page 3 Examination and Preparation for Use We are delighted to have you as a Supera Anesthesia Innovations customer and want you to be completely satisfied with your purchase. Please inspect the contents of your order to see if everything is as you expected. Should anything not be exactly right or if anything was damaged in shipping, please contact your sales representative right away for help.
  • Page 5 OC8200 Machine Assembly * NOTE: THIS REQUIRES TWO PEOPLE TO SAFELY ASSEMBLY * 1. Carefully unpack the top assembly and frame from the box 2. Align the top to the frame as shown. NOTE: The frame base has two cross bars with holes in them for mounting the concentrator.
  • Page 6 1. Remove the oxygen concentrator from the box and guide the 4 threaded post into the frame 2. Install and tighten the nuts. DO NOT OVER TIGHTEN THE NUTS, SNUG FIT IS ALL THAT IS NEEDED...
  • Page 7 Oxygen Hose Setup Attach the green hose from your anesthesia machine to one of the threaded fittings. NOTE: Both output the same DO NOT OVER TIGHTEN Attach the green hose from the OC8000 top assembly to the threaded fitting on the concentrator. DO NOT OVER TIGHTEN...
  • Page 8 Auxiliary & Main Oxygen Switch MAIN Switch Position With the switch in the “MAIN” position the supply of oxygen to the anesthesia machine comes directly from the oxygen concentrator. NOTE: It does not fill the “E” tank. AUXILARY Switch Position With the switch in the “AUX”...
  • Page 9: Oxygen Concentrator

    USER'S GUIDE OC8000 OXYGEN CONCENTRATOR FOR VETERINARY USE ONLY GLOSSARY OF SYMBOLS : ON (power switched on) : Do not use oil or grease : OFF (power switched off) : Technical information : Consult the accompanying documents : Keep in the vertical position : Class II protection : Fragile - handle with care : Do not expose to open flames...
  • Page 10: General Safety Guidelines

    GENERAL SAFETY GUIDELINES Only persons who have read and understood this entire manual should be allowed to operate the O2 concentr The CAUTIONS below indicate a potentially The WARNINGS below indicate a potential hazardous situation. If conditions are not hazardous situation. If conditions are not avoided a situation could occur that results avoided a situation could occur that results in property damage or minor injury or both.
  • Page 11: Method For Disposing Of Device

    CONFORMITY WITH IEC60601-1 (2nd 1.1 METHOD FOR WASTE DISPOSAL Edition) All waste from the device (Patient Circut, "The manufacturer, assembler, installer or Filters, Etc.) must be disposed of using methods distributor are not considered to be responsible appropriate to the civil authority of the location themselves for the consequences on the safety, where disposed.
  • Page 12 2.1. Front panel (Fig. 2.1) I/O (ON/OFF) switch 2 - Indicator lights 3 - Oxygen product outlet 4 - Circuit breaker 5 - Flow meter...
  • Page 13: Filter Maintenance

    3.2 Turning off device 3.1 Turning on device At the end of the usage, press the I/O Switch a. Plug the power cable into a power outlet to place it in the O (OFF) position to stop the (Fig. 3.3) of the correct voltage and frequency device.
  • Page 14: Useful Information

    4.3. Maintenance NO INTERNAL MAINTENANCE 5.2. Materials in direct or indirect contact with the product output IS REQUIRED OR SHOULD BE PERFORMED. Concentrator casing ........ABS Power Cord ........... PVC Cabinet Air Filter ......Polyester OPENING THE CASE WILL I/0 (On/Off) switch........Nylon Casters............Nylon VOID THE WARRANTY.
  • Page 15 Process fault 5.4.3 Indicators (continued) • In the case of a process fault, a visible and au- • The red indicator warns of a process fault. dible alarm is activated (continuous red light or One event that can cause the red indicator lighted alarm and audible alarm).
  • Page 16 5.5.4 Maintenance of the Device Alarms 5. 5. OCSI (oxygen concentration status indication module) function No special maintenance is required. The alarm set-point is factory set and the setting cannot be 5.5.1. Operating principle adjusted. All models are set at 84%. The oxygen monitor (Item 2 Fig 2.1) is an The equipment supplier verifies that the device electronic module capable of checking the...
  • Page 17: Air Circulation

    One tubeaxial fan cools the compressor compart- ment and a second fan cools the heat exchanger coil. Use original parts only Environmental limit conditions Contact Supera Anesthesia Innovations The performances of the device (especially the for additiional information oxygen concentration) are quoted at 21 C (70 and one atmosphere.
  • Page 18: Troubleshooting

    5.8. Troubleshooting. Observations Possible Causes Solutions The I/O (ON/OFF) button is in the "I" Power cord is not correctly (ON) position but the device does not plugged into the wall outlet. Check the cable connection. operate. Check the circuit breaker (5) on the The audible alarm sounds intermit- Power failure.
  • Page 21: Maintenance

    P/N OC8000-2 Fine inlet filter (square black plastic filter) The fine particle filter goes under the course filter. This is a replacement only filter. (non-washable) Failure to monitor the condition of the filters will void the warranty. Contact Supera for replacement 877-620-1500...
  • Page 23: Warranty / Return Policy

    The Supera LLC Limited Warranty covers parts and construction of all products manufactured by Supera LLC in its USA manufacturing facility. If we find that a Supera machine does not operate as stated in the instructions and specifications that come with the product due to defects in design, materials, or workmanship, Supera will repair the unit or replace it with one of equal or greater value at no charge to the customer for up to three years from the date of purchase.
  • Page 25 DOCUMENTATION Date purchased: Purchased from: Machine serial number: Machine service information Service Date: Service information: Service Date: Service information: Service Date: Service information:...
  • Page 26 CS@superavet.com 7. Note: unless the following procedures are followed correctly and we are notified within ten (10) days, SUPERA LLC cannot accept any responsibility for the problems that may ensue. DO NOT RETURN ANY DAMAGED GOODS TO SUPERA LLC WITHOUT PRIOR AUTHORIZATION OF SUPERA LLC AND THE CARRIER.
  • Page 28 (503) 723-5068 or (877) 620-1500 toll-free E-Mail: info@superavet.com 12330 S.E. Highway 212, Clackamas, OR 97015 www.SuperaVet.com...

This manual is also suitable for:

Pureline oc8200Pureline m8000