Calibration
Special Applications
• When the dielectric constant ε
e.g. with plastic granulate, it may be necessary to change the factory-set
value.
Calibration is possible for ε
• A change in the factory-set value may also be required if the minimum
distances cannot be maintained.
For this calibration, the silo must be empty or the level must be at least
100 mm below the probe.
The adjuster for calibration is approx. 30 turns from one end of the range to
the other.
Carry out the calibration according to the sequence of diagrams Fig. 17 to
Fig. 19.
of the material is smaller than 2.0,
r
> 1.6.
r
Capacitance Calibration
(Changing Factory-Set
Adjustment)
Capacitance Calibration,
Initial Settings
Fig. 17
These initial settings must be done before the
capacitance calibration.
21