Remove the old Exhaust-Silencer (if necessary) ..............6 Fit the new Exhaust-Silencer ..................9 3.5.1 Fit the new Exhaust-Silencer to a QDP, iQDP or CDP i, Ei or Si pump ........... 9 3.5.2 Fit the new Exhaust-Silencer to a CDP Ni pump ..............10 Leak-test the installation ....................14...
Page 4
QDP pump services ...................... 8 Correct mounting clamp orientation (CDP pump shown) ............10 Installation of the Exhaust-Silencer on CDP i, Ei and Si pumps ..........11 Installation of the Exhaust-Silencer on QDP and iQDP pumps ...........12 Installation of the Exhaust-Silencer on CDP Ni ..............13...
Description Two models of Exhaust-Silencer are available, one model for use with the Edwards CDP range of pumps and one model for use with the Edwards QDP and iQDP ranges of pumps. The two models are identical, except that the Exhaust- Silencer for the CDP pumps has mounting hooks (Figure 1, item 14).
System requirements WARNING Use bellows which have a suitable pressure rating. If the exhaust pipeline is blocked, the CDP, QDP and iQDP pumps can generate pressures of up to 7 bar in the pipeline. WARNING Ensure that the pressure in the exhaust pipeline does not exceed 6 psi.
1. Release the NW40 clamp which secures the inlet-flange of the Exhaust-Silencer to the CDP, QDP or iQDP pump outlet or the outlet of the long NW40 elbow connected to the outlet of the CDP, QDP or iQDP pump. Remove the clamp and the Co-Seal.
3.5.1 Fit the new Exhaust-Silencer to a QDP, iQDP or CDP i, Ei or Si pump Refer to Figures 6 and 7 to determine the correct installation configuration for your pump and fit the new Exhaust- Silencer as described below.
A386-11-880 Issue E Figure 5 - Correct mounting clamp orientation (CDP pump shown) 3.5.2 Fit the new Exhaust-Silencer to a CDP Ni pump The correct installation configuration for your pump is shown in Figure 8. Fit the new Exhaust-Silencer as described below.
Page 15
Figure 6 - Installation of the Exhaust-Silencer on CDP i, Ei and Si pumps 1. Exhaust check-valve 2. Exhaust-Silencer 3. Mounting clamp...
Page 16
Figure 7 - Installation of the Exhaust-Silencer on QDP and iQDP pumps 1. Exhaust check-valve 2. Clamp 3. Exhaust-Silencer 4. Alternative outlet position 5. Mounting clamp...
Page 17
Figure 8 - Installation of the Exhaust-Silencer on CDP Ni 1. Exhaust check-valve 2. Exhaust-Silencer 3. Mounting clamp...
Fluorinated materials are used in the Exhaust-Silencer and the CDP, QDP and iQDP pumps. Fluorinated materials are safe in normal use but can decompose into very dangerous substances (which may include hydrofluoric acid) if they are heated to 260 ºC and above.
10. Leak-test the silencer. 11. Refit the Exhaust-Silencer to the pump: For QDP, iQDP and CDP i, Ei and Si pumps, refer to Steps 6 to 9 of Section 3.5.1. For CDP Ni pumps, refer to Steps 3 and 4 of Section 3.5.2.
The majority of these centres employ Service Engineers who have undergone comprehensive Edwards training courses. Order spare parts and accessories from your nearest Edwards company or distributor. When you order, please state for each part required: Model and Item Number of your equipment ...