219 Route 206
PO Box 372
Andover, NJ 07821 USA
Note: When using analog outputs, there is an additional ±1% error in the
analog output voltage. This is in addition to the normal measurement error.
For example, an air velocity reading of 500 FPM would normally have an
accuracy of ±1% of reading ±1 digit (±6 FPM) when the data is viewed
on the LCD display.
With the additional error associated with the analog output voltage, the
effective accuracy of the analog output for this air velocity measurement
will be ±2% of reading ±1 digit (±11 FPM).
Custom Analog Voltage Outputs are also available – contact Kanomax for
details.
Analog Output circuit block diagram and connector pin assignment
6824
User Manual, PN 10283
Rev 1.0, 30-July-2021
Copyright © 2021, Kanomax USA, Inc.
TM
Phone: 1-800-247-8887
Fax:
(973) 786-7586
info@kanomax-usa.com
Mating Cable Connector:
Binder Part No. 99-0413-00-05
Kanomax USA, Inc