5
Calibrating the Recorder
5a. If the first character is correct,
5b. If the first character is not correct,
6.
7a. If the second character of the humidity calibration code is correct,
7b. If the second character is not correct,
5-2
A 2-character humidity calibration code should now appear in the display
with the first character blinking. Refer to Figure 5-2. Remember, this
calibration code must match the code on the sensor probe.
Figure 5-2. Humidity Code - First Character Blinking
press the "SET" button and go to step 6.
press the "HIGH" button to increase the value of the character or the
"LOW" button to decrease the value of the character. When this character
is correct, press the "SET" button.
The second character should now be blinking. Refer to Figure 5-3. If it is not,
press the "SET" button until it does.
Figure 5-3. Humidity Code - Second Character Blinking
press the "SET" button and go to step 8.
press the "HIGH" button to increase the value of the character or the
"LOW" button to decrease the value of the character. When this character
is correct, press the "SET" button.
RH
57
°C
°F
RH
°C
57
°F