Omega CT485B-110V-G-AL User Manual page 16

High performance microprocessor-based temperature and humidity recorder
Table of Contents

Advertisement

To change the chart paper:
1a. Table Mounted Recorder:
To balance the recorder, fully extend the stabilizing arm at the bottom of the
recorder. Refer to Figure 2-8. Continue to Step 2.
1b. Wall Mounted Recorder:
You can also leave the recorder on the wall if it is easily accessible.
2.
Open the door and hold it open with left hand.
3.
Remove the magnetic chart paper knob (Figure 2-9, item #2) and any used
chart paper.
4.
Place the new chart paper (Figure 2-9, item #1) on the spindle and rotate it
until the current time on the chart paper is aligned with the time arrow on the
chart base of the recorder. See Figure 2-9 detail.
5.
When the chart paper is correctly oriented, replace the magnetic knob
(Figure 2-9, item #2), being careful not to alter the position of the chart paper.
6.
Make sure that the "Chart Type" switch, located behind the control panel door,
is in the correct position relative to the chart paper used.
If the switch is in the wrong position, the chart paper will not
rotate at the correct speed causing the chart time not to be
synchronized with the current time.
Figure 2-8. Stabilizing Arm Fully Extended
NOTE
Setting Up the Recorder
H
U
M
I
D
I
T
Y
2
2-7

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ct485b-220v-g-alCt485b-110v-w-alCt485b-220v-w-al