Removing The M9035A From The Chassis; Power On Sequence; Power Off Sequence - Keysight Technologies M9035A Startup Manual

Pxie embedded controller
Table of Contents

Advertisement

Step 3: Install the M9035A Embedded Controller
9
10

Removing the M9035A from the Chassis

To remove the M9035A controller, reverse the previous procedure. Shut down
the operating system to power-down the chassis. Loosen the four captive
screws, press down on the ejector/injector handle and pull outwards to release
the module from the backplane. Pull the module towards you until it is free of the
chassis.

Power On Sequence

The M9035A controller, like any PC, will boot with the installed operating system.
Refer to the M9035A User Guide for detailed power on and configuration
sequence information.

Power Off Sequence

As with any PC, Windows-based instruments should not be shutdown by either
turning off the power via an external power source or by pulling the power plug
out from the chassis rear panel. This could corrupt the operating system. One
method to shut down the controller is to execute the Windows shutdown process
which shuts down the M9035A controller then turns off the PXIe chassis. A
second method, if the chassis rear panel INHIBIT switch is set to DEF, is to
momentarily press the front panel button. Either method shuts down the
M9035A safely.
Keysight M9035A PXIe Embedded Controller Startup Guide
Connect other peripherals (such as the keyboard, mouse, monitor, LAN cable,
etc.) to the Embedded Controller.
Power up the PXIe chassis. Verify that the chassis fans are operating and free
of obstructions that may restrict air flow.
Depending on the system configuration, the BIOS may take 90
seconds (or longer) to boot. During this time, the Keysight splash
screen (the Keysight logo) is present and it may appear that
nothing is happening. This is generally caused by the inclusion of
certain devices in the switch fabric (such as a RAID controller) that
require a longer time for PCI enumeration.
If this product is not used as specified, the protection provided by
the equipment could be impaired. This product must be used in a
normal condition (in which all means for protection are intact) only.
17