Scale Error Messages - Seca 334 LC Service Manual

Baby scale with 4 load cell technology
Table of Contents

Advertisement

Troubleshooting
3.2

Scale error messages

The scales have an error message feature which indicates possible malfunctions and
defects. These error messages are described below.
3.3
Error messages
[334|335|382|383|384|385|833|834|835|874|875|876|877|878|952|956]
If a malfunction or defect occurs during operation of the scale, it is shown on the display
as an error message. Error messages are displayed in the following form:
"E xx"
E:
Error
xx:
Error number (see Table 4)
Displayed
Load cell defective, supply
E 11
leads damaged or electronics
defective
Preload too great, load cells
E 12
damaged or electronics
defective
Limit for automatic zero tracking
E 13
reached
Error when determining zero
E 16
point on switch-on
03.04.2013/JEG
Cause
16 (24)
Remedy
Diagnosis with seca serva:
Check load cell connections
Poss. replace load cell or
electronics
Check preload on scale
Diagnosis with seca serva:
Poss. replace load cell or
electronics
Place load of >0.5d/s on scale
Protect scale from vibration /
shocks during start-up
seca
30-34-00-864