Troubleshooting
3.2
Scale error messages
The scales have an error message feature which indicates possible malfunctions and
defects. These error messages are described below.
3.3
Error messages
[334|335|382|383|384|385|833|834|835|874|875|876|877|878|952|956]
If a malfunction or defect occurs during operation of the scale, it is shown on the display
as an error message. Error messages are displayed in the following form:
"E xx"
E:
Error
xx:
Error number (see Table 4)
Displayed
Load cell defective, supply
E 11
leads damaged or electronics
defective
Preload too great, load cells
E 12
damaged or electronics
defective
Limit for automatic zero tracking
E 13
reached
Error when determining zero
E 16
point on switch-on
03.04.2013/JEG
Cause
16 (24)
Remedy
Diagnosis with seca serva:
Check load cell connections
Poss. replace load cell or
electronics
Check preload on scale
Diagnosis with seca serva:
Poss. replace load cell or
electronics
Place load of >0.5d/s on scale
Protect scale from vibration /
shocks during start-up
seca
30-34-00-864