Linearity Adjustment; Connections And Initial Warm Up; Zero Adjustment; 25% Flow Adjustment (Using R33 Potentiometer) - Omega FMA1700A User Manual

Mass flowmeter
Hide thumbs Also See for FMA1700A:
Table of Contents

Advertisement

See also section 8. TROUBLESHOOTING. If after attempting to remedy the
above conditions, a malfunction still persists, return the meter for factory
service, see section 1.1.
At this point the calibration is complete. However, it is advisable that several
additional points between 0 and 100%, such as 25%, 50%, and 75% flow be
checked. If discrepancies are found, proceed to step 7.3 for Linearity Adjustment.
7.3

Linearity Adjustment

7.3.1

Connections and Initial Warm Up

At the 9-pin "D" connector of the FMA1700A & FMA1800A transducer, connect
the multimeter to output pins [2] and [3] for 0-5 VDC (or pins [8] and [9] for 4-20
mA)-(see Figure 2.a). If calibration to a new flow range or different gas is being
performed, it may be necessary to remove any jumpers at J1.A, J1.B, J1.C, and
J1.D before the beginning the linearizing procedure.
Power up the Mass Flow Meter for at least 30 minutes prior to commencing the
calibration procedure.

7.3.2 ZERO Adjustment

Shut off the flow of gas into the Mass Flow Meter. To ensure that no seepage or
leak occurs into the meter, it is good practice to temporarily disconnect the gas
source. Using the multimeter and the insulated screwdriver, adjust the ZERO
potentiometer [R34] through the access window for 0 VDC (or 4 mA respectively)
at zero flow.
CAUTION: The minimum voltage on 0-5 Vdc output can be in the range of 7
to 25 mV. Trying to reduce voltage below this level may increase negative
zero shift. This shift may be invisible on devices without LCD display. Stop
R34 zero potentiometer adjustment if voltage on 0-5 Vdc output is in the
range from 7 to 25 mV and does not decrease any lower.

7.3.3 25% Flow Adjustment (using R33 potentiometer)

Reconnect the gas source. Using the flow regulator, adjust the flow rate to 25%
of full scale flow. Check the flow rate indicated against the flow calibrator.
Adjust the setting for potentiometer [R33] by using the insulated screwdriver
through the access window, until the output of the flow meter reads 1.25VDC
±63mV (or 8mA ±0.25mA).
Using the flow regulator, adjust the flow rate until the output of the flow meter
reads 0.5 VDC (or 5.6mA). Check the flow rate against the flow calibrator. If the
flow rate indicated by the calibrator is within 10% ± 1.5% of F.S. then skip para-
graphs 7.3.4, 7.3.5 and proceed directly to paragraph 7.3.6, if not, perform 10%
flow adjustment according to paragraph 7.3.4.
15

Advertisement

Table of Contents
loading

This manual is also suitable for:

Fma1800aFma1716a-st-c

Table of Contents