Commissioning
Menu item
Temp. adjustment
(only with Measuring
mode: Cl (mg/l))
Conc. offset
Save and quit
Quit
28
Selection/Values
-1.5 °C ... +1.5 °C
-10.0 mg/l ... +10. mg/l
Explanations
The temperature compensation function
enables the temperature sensor to be
balanced against a reference temperature
measurement (displacement of the zero
point by ±1.5 °C).
Notes:
 Due to the thermal capacity of the sen-
sor, it is necessary to place it in a con-
tainer with at least 2 liters of water.
 Leave the sensor in this container for at
least 15 minutes while stirring occa-
sionally, then carry out the adjustment.
If the temperature difference of the
water and sensor is > 10°C, leave the
sensor in the container for at least one
hour while stirring occasionally.
Constant value that is added to the mea-
sured value.
The system confirms the saving of the set-
tings and the display switches to the next
higher level.
The display switches to the next higher
level without storing the new settings.