Cleaning accessories
The MicroRPM unit is protected from contamination by the
MicroGard IIC Filter kit (V-892392 MicroGard IIC Filter, disposable
mouthpiece/nose clip) during mouth pressures measurements.
The MicroGard II Filter Adapter for RPM (36-ASS1307), the
Expiratory and Inspiratory Pressure Valve Assembly (36-ASS1221,
36-ASS1222) are reusable parts and needs to be disposed after 6
months of usage. The Nasal Probes (36-NPROBE01, 36-NPROBE02,
36-NPROBE03, 36- NPROBE04), however, may be immersed in a
cold sterilising solution such as PeraSafe (36-SSC5000A). Rinse
thoroughly and leave to dry before reassembly.
Important note: Nasal Probes, which are not sterilized, must be
disposed off immediately after each use.
If there are changes on the material surfaces (cracks, brittleness)
the respective parts must be disposed off.