Cleaning Instructions; External Surfaces Of The Spirometer; Cleaning Accessories - Vyaire MicroRPM Operating Manual

Respiratory pressure meter
Table of Contents

Advertisement

Cleaning Instructions

Disinfection of contaminated parts is only effective after having
them carefully pre-cleaned. Vyaire recommends the tested solution
of PeraSafe sterilizing powder (36-SSC5000A) for pre-cleaning and
disinfection. If a different solution is used, please follow the given
manufacturer's instructions.
The device must not be wiped with any aqueous solutions and must
not be exposed to solvents e.g. alcohol, chloride solutions as there
are electronic components inside that will be permanently damaged.
CAUTION: Switch off the device before cleaning.

External Surfaces of the Spirometer

CAUTION: Do not attempt to wash or immerse the Respiratory
Pressure Meter in water or cleaning fluid, as there are electronic
components inside that will be permanently damaged
The external housing of the pressure meter may be wiped with
sterile wipes or a damp cloth that has been immersed in a cold
sterilising solution.

Cleaning accessories

Cleaning accessories The MicroRPM unit is protected from
contamination by the MicroGard IIC Filter kit (V-892392 MicroGard
IIC Filter, disposable mouthpiece/nose clip) during mouth
pressures measurements. The MicroGard II Filter Adapter for
RPM (36-ASS1307), the Expiratory and Inspiratory Pressure Valve
Assembly (36-ASS1221, 36-ASS1222) are reusable parts and
needs to be disposed after 6 months of usage. The Nasal Probes
(36-NPROBE01, 36-NPROBE02, 36-NPROBE03, 36- NPROBE04),
however, may be immersed in a cold sterilising solution such as
18

Hide quick links:

Advertisement

Table of Contents

Troubleshooting

loading
Need help?

Need help?

Do you have a question about the MicroRPM and is the answer not in the manual?

Subscribe to Our Youtube Channel

Table of Contents

Save PDF