CAUTION: The operator should not touch the contacts of the
battery and the patient at the same time.
Please Note: Dispose of the waste battery in accordance with EU
Waste Battery Regulations.
Cleaning Instructions
Disinfection of contaminated parts is only effective after having
them carefully pre-cleaned. Vyaire recommends the tested solution
of PeraSafe sterilizing powder (36-SSC5000A) for pre-cleaning and
disinfection. If a different solution is used, please follow the given
manufacturer's instructions.
The device must not be wiped with any aqueous solutions and must
not be exposed to solvents e.g. alcohol, chloride solutions as there
are electronic components inside that will be permanently damaged.
CAUTION: Switch off the device before cleaning.
External Surfaces of the Spirometer
CAUTION: Do not attempt to wash or immerse the Respiratory
Pressure Meter in water or cleaning fluid, as there are electronic
components inside that will be permanently damaged
The external housing of the pressure meter may be wiped with
sterile wipes or a damp cloth that has been immersed in a cold
sterilizing solution.
43