Battery; Battery Replacement - Vyaire MicroRPM Operating Manual

Respiratory pressure meter
Table of Contents

Advertisement

Battery

The battery level is automatically checked when the unit is switched
on.
When the battery is low 'bAt' will be displayed before the auto-zero
procedure. The MicroRPM may be used when this occurs, but the
battery should be replaced as soon as possible.
When the battery is completely exhausted the unit will beep twice
and turn itself off immediately.

Battery Replacement

Locate the sliding cover situated on the rear of the unit, towards the
bottom of the device.
Place your thumb over the round thumb indent, press gently and
slide the cover to the right to remove it from the unit.
Lift the old battery out, and holding the battery terminal by the
plastic body, pull it off the old battery.
Plug the new battery into the battery terminal, taking care that the
correct polarity is observed.
Push the battery back into the battery holder, and replace the
battery cover onto the guides. Slide the battery cover to the left
until it is fully home.
Note: Please remove the battery if the meter is likely to be unused
for some time.
CAUTION: Do not open the battery cover when the device is
turned on.
CAUTION: The operator should not touch the contacts of the
battery and the patient at the same time.
Please Note: Dispose of the waste battery in accordance with EU
Waste Battery Regulations.