Testing - DELHI 90
A test ring is needed to ob-
tain good sealing during
the HI test. Place the ring
on the 2cc coupler tube
and push it up as shown
1
2
6
Not subject to issue control when printed
Important points to obtain a good & stable HI test:
1.
Test ring to make good sealing
2.
HI to be placed in a straight line
3.
HI almost floating, only touch the foam slightly
4.
HI prog adapt must not touch test jig
5.
Prog./Batt. cables organized, must provide less
driven force to HI
2
6.
Coupler cable organized, in order to avoid any
change of coupler position when closing sound
chamber
1
4
5
3
Doc 0213850 rev M
Page 22 of 22