Measuring capacitance:

1) Set the function switch to the
2) Connect the adapter to the "
3) If the reading on the screen is not zero, press the "REL" button and the meter will enters the relative value measurement mode,
and then the "REL" symbol appears on the LCD and the reading becomes zero.
4) Short the two pins of the capacitor to be measured and fully discharge the capacitor, and insert the capacitor into the capacitor
jack on the adapter.
5) Take the readings indicated on the screen after it is stable.
Tips:
When the capacitance measurement function is activated, ranges are switched automatically. Since the capacitance is
calculated using the charge/discharge cycle of the measured capacitor, the larger the measured capacitance, the more time
the measurement takes.
Measuring temperature (only applies to 03017):
1) Set the function switch to position "
2) Insert the negative plug at the cold junction of the Type K thermocouple into "
"
"jack.
3) Place the other end of the thermocouple at the place where the temperature measurement will occur.
4) Take the reading of temperature indicated on the screen.。
Note:
To avoid damage to the multimeter or other equipment, remember that the rated range of multimeter is -20-1000° C
(-4-1832° F). The Type K thermocouple supplied with the multimeter as a giveaway has the rating of 250℃ , which
is for reference only. It is a nonprofessional grade product. To obtain accurately measured temperature, please use a
professional grade thermocouple.
03015/03017
position.
" jacket and "
". And press "
中文
EN
" jacket as shown.
" button to select the unit of temperature: ℃ / ℉ .
" jack, and the positive plug into