Installing the Transmittance Zero Calibration Plate (Option)

The transmittance zero calibration plate is used when performing 0% calibration for transmittance measurement.
• When using the optional SpectraMagic NX, set the measurement method to transmittance in advance.
• Install the target mask for LAV measurement and the white calibration plate onto the instrument.
Procedure
1 Open the transmittance specimen chamber cover.
2 Install the transmittance zero calibration plate
in a position so that it completely covers the
illumination window.
Memo
• When using the optional transmittance specimen
holder, install the transmittance zero calibration
plate into the transmittance sample holder. For
information about installing the transmittance
specimen holder, refer to the instruction manual
that was provided with the transmittance specimen
holder.
3 Close the transmittance specimen chamber cover.
Precautions for Use of the Transmittance Zero Calibration Plate
• Be careful that the transmittance zero calibration plate does not become scratched and does not contact
fingerprints or other dirt.
• If the transmittance zero calibration plate becomes dirty, wipe it gently with a soft, clean, and dry cloth.
• If the dirt does not come off easily, wipe using a soft cloth dampened with ethanol.
• If the transmittance zero calibration plate is scratched or the dirt cannot be removed, replace it.
25
Illumination
window