PRS 500 X Basic diagnostic X-ray system Model/ID: 7066-9-8050 Instructions for use Ident. Nr. 5066-0-8002 PRS 500 X in analogue base configuration Version: 5.0 PROTEC GmbH & Co. KG Issued: 2020-11-12 In den Dorfwiesen 14, 71720 Oberstenfeld, Germany Subject to alterations...
Page 2
All sheets of this document contain proprietary and confidential information of PROTEC GmbH & Co. KG and is intended for exclusive use by current PROTEC GmbH & Co. KG customers. Copying, disclosure to other or other use is prohibited without the express written authorization of PROTEC´s law department.
Setting the X-ray unit on the mid moving Bucky, Grid entity ............22 4.1.1.3 Inserting a cassette into the cassette tray ....................22 4.1.1.4 Adjusting the focus-film distance (FFD) ......................22 PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 3 von 40...
Page 4
Description of symbols, labels and abbreviations ............36 Symbols ....................................... 36 Identification label..................................37 Labels ........................................38 Position symbols and labels ..............................39 Abbreviations ....................................40 PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 4 von 40...
X-ray radiation. The intensity depends on the adjusted values of voltage, current and time. The radiation comes orthogonal out of the X-ray tube and is limited by the collimator. PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 6 von 40...
PROTEC GmbH & Co. KG Customers. Assemblers and other Customers not employed by nor directly affiliated with PROTEC GmbH & Co. KG technical services are directed to contact the local PROTEC GmbH & Co. KG office before attempting installation or service procedures.
This user manual describes the special features and operational aspects of the PRS 500 X, knowledge of which are required for efficient and effective use of the radiographic system. Prior to working with the PRS 500 X, it is required that the user reads the Safety Notes as well as the chapter regarding operation.
The radiographic system is primarily made of metal pieces. This has a main role in the weight of the device. The radiographic system PRS 500 X has a weight of 642kg (incl. Generator), option: telescopic arm 30kg (additionally). Every technician is obliged to check the ground load. Also double bottoms and hollow floors have to be taken into account.
Intended user group The radiographic system PRS 500 X is exclusively designated for use by professional who are trained, in accordance with the corresponding national regulations, in the use of diagnostic X-Ry equipment and its proper intended use in connection with other medical devices, objects and accessories.
Adjustments and calibrations that are described within the user manual must be made, with the aid of The technical description for the system, by the PROTEC GmbH & Co. KG customer service department or a PROTEC GmbH & Co. KG authorized service technician.
Page 13
It’s not allowed to make any medical not indicated exposures on people. At pregnancy or children the question is if the exposure is really necessary. If possible it’s better to abandon it. PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 13 von 40...
Duty Cycle: S3 15% - maximum continuous operation of 1,5 minutes. 2.1.4 Operating personnel The radiographic system PRS 500 X with stand should only be operated by personnel who are trained in accordance with the corresponding national regulations in the use and operation of diagnostic radiographic systems.
2.1.6 Explosion protection These radiographic system PRS 500 X is not designated for use within areas with explosive hazards. 2.1.7 Radiation protection X-ray radiation can pose a hazard to patients and other people when the regulations regarding the operation of radiographic systems are not followed.
The user must take remedial measures such as implementation or reorientation of the device. The PRS 500 X is intended for the usage in a professional environment of the medical service (e.g. clinic, surgery centers, physiology offices …) PROTEC GmbH &...
In this state no x-rays may be taken. The X-ray tube carrier must first engage in one of the positions (0 / ± 90° / ± 180°)! PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 17 von 40...
Page 18
If the operating unit is equipped with handles, the electromagnetic lock of one or more movements can be released by thumb pressure on the keys of the operating unit and the tube head can be brought into the desired position. PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 18 von 40...
Page 19
There is an increased risk of injury if the X-ray tube carrier is not engaged! Detailed information please find in the accompanying User Manual PROGNOST SH. PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 19 von 40...
Detailed information please find in the enclosed User Manual of the X-ray generator. Control elements of Bucky, Grid entity Detailed information please find in the enclosed User Manual. PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 20 von 40...
Detailed information please find in the enclosed User Manual of the RAPIXX system. Control elements and device displays of CONAXX 2 Detailed information please find in the enclosed User Manual of the CONAXX 2. PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 21 von 40...
Start the exposure by pressing the switches for exposure preparation/exposure release. 4.1.1.7 Overtable exposures • Place a cassette to the desired position on the table top. PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 22 von 40...
Page 23
Push the tube column to the lateral detector holder − Adjust the height of the tube arm or of the table that the light field is aligned to the panel PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 23 von 40...
(Exposure every 30 seconds, otherwise tube may arc). See User Manual Generator and CONAXX 2 User Manual chapter 5.3. Detailed information please find in the enclosed User Manual of the X-ray tube. PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 24 von 40...
4.9.1 Switching On/Off the PRS 500 X Switching on the PRS 500 X happens via the control panel of the Generator. The Generator supplies every system component with power. On the Generator and the control panel will run a self-test when switching them on. After the self-test was successful the parameters will be displayed which can be saved under Organ-number #0.
4.10 Exposure automatic If the PRS 500 X is operated with an exposure automatic the functionality can be checked like this: Place a Phantom or any other weakening object in the radiation way. Choose a measuring chamber and expose.
Cleaning The use of corrosive or abrasive cleaning agents as well as solvents is not allowed. These materials can cause damage to the outer surface of the radiographic system PRS 500 X or to the coating of the individual components.
5.3.5 Product life time The PRS 500 X has an expected product life of 10 years when used in accordance with the product specifications/ limitations and provided that maintenance through the PROTEC service department or a PROTEC authorized service provider has be completed. After reaching the life span the further usage of the device happens on own risk.
5.3.8 Disposal The X-ray system PRS 500 X contains different plastics, oils and heavy metal. At disposal of exchange parts or the whole system the current regulations have to be observed. Please contact the service or a company specialized for disposing the components.
Mobile HF-Communication devices (radios) (including their accessories such as antenna cables and external antennas) shouldn’t be used closer than 30cm (12 Inch) to the marked parts and cables of the PRS 500 X. Disregarding this can cause a decrease in the performance features of the device.
PRS 500 X 5066-0-8002 Instructions for use Technical Data Dimensions PATIENT PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 32 von 40...
Page 33
PRS 500 X 5066-0-8002 Instructions for use PATIENT *OP – Operator PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 33 von 40...
< 0,5 Al mm according to EN 60601-1-3. Tested at 100 kV with a first half-value layer thickness (HVLT) of 3,7 mm Al. 7.4.1 Protection Art and Protection Class The PRS 500 X is consistent with a protection class 1 device and contains applicable parts Type B (according to EN 60601-1).Environmental conditions. Automatic cutoff dose 7.5.1 Analogue System The automatic cutoff dose is 2.5 µGy.
Environmental Conditions for Shipping and Storage Ambient Temperature - 10°C to + 70°C Relative humidity 10% to 95% (non-condensing) Atmospheric pressure 500 hPa to 1060hPa PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 35 von 40...
Homepage www.protec-med.com/download Notes on disposal; WEEE , Waste of Electrical and Electronic Equipment Protective ground (Earth) Caution: pinch-/crushing hazard for hands and fingers PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 36 von 40...
Horizontal movement of the X-ray system tube support, floor stand and vertical movement of the X-ray tube assembly Identification label Exemplary for PRS 500 X with X-ray generator 50 kW PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 37 von 40...
Caution: Possible pinch-/crushing hazard fort he hands and fingers while moving the X-ray tube assembly unit. Labels on the X-ray system tube support, floor stand and column X-ray system image receptor stand Product label PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 38 von 40...
5066-0-8002 PRS 500 X Instructions for use Position symbols and labels At both ends of the rail Identification label Identification label PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 39 von 40...
Instructions for use Abbreviations Abbreviations Millimetres Centimetres Pound Kilogram °C Degree -Celsius Hectopascal German Industry Standard European Standard CE-Mark Hertz Duty cycle Ampere Serial number PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 40 von 40...