Protec PRS 500 F Instructions For Use Manual

Protec PRS 500 F Instructions For Use Manual

Basic diagnostic x-ray system
Hide thumbs Also See for PRS 500 F:
Table of Contents

Advertisement

Quick Links

PRS 500 F
Basic diagnostic X-ray system
Model/ID: 7067-9-8050_Vxxx
Basis_UDI-DI: 426050264X006ZM
Instructions for use
Ident. Nr. 5067-0-8002
PRS 500 F in analogue base configuration
Version: 6.0
PROTEC GmbH & Co. KG
Issued: 2022-01-19
In den Dorfwiesen 14, 71720 Oberstenfeld, Germany
Subject to alterations
Telephone: +49 (0) 7062/92 55-0 E-Mail: protec@protec-med.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PRS 500 F and is the answer not in the manual?

Questions and answers

Summary of Contents for Protec PRS 500 F

  • Page 1 Basic diagnostic X-ray system Model/ID: 7067-9-8050_Vxxx Basis_UDI-DI: 426050264X006ZM Instructions for use Ident. Nr. 5067-0-8002 PRS 500 F in analogue base configuration Version: 6.0 PROTEC GmbH & Co. KG Issued: 2022-01-19 In den Dorfwiesen 14, 71720 Oberstenfeld, Germany Subject to alterations...
  • Page 2 All sheets of this document contain proprietary and confidential information of PROTEC GmbH & Co. KG and is intended for exclusive use by current PROTEC GmbH & Co. KG customers. Copying, disclosure to others or other use is prohibited without the express written consent of PROTEC´s law department.
  • Page 3: Table Of Contents

    Control elements of Bucky, Grid entity ..........................22 Control elements and Display of verticalX-ray system image receptor stand PROVERT ....23 3.9.1 Vertical carriage .................................. 23 PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 3 von 44...
  • Page 4 X-ray system table PROGNOST F ............................37 Bucky, Grid entity ................................... 37 X-ray Tube Support Stand ................................ 38 Image Receptor Stand ................................38 Attenuation Equivalent ................................39 PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 4 von 44...
  • Page 5 Description of symbols, labels and abbreviations ............40 Symbols ....................................... 40 Identification label..................................41 Labels ........................................42 Position symbols and labels ..............................43 Abbreviations ....................................44 PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 5 von 44...
  • Page 6: Revision Status

    1.2.2; Kap. 1.2.2.1; Kap. Rotation; Note Installation 1.3.2; Kap. 7.1 new; total weight changed; ceiling height changed; Traverse paths PROGNOST SH changed vertically; translation cpl. revised PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 6 von 44...
  • Page 7: General Notes

    The X-ray components from PROTEC used are only devices for the human medical diagnostic area, which can be set up to a maximum of 150 kV. Further information can be found in the technical data inside the instructions for use for the generators, X-ray tubes and collimators.
  • Page 8 PROTEC GmbH & Co. KG. Assembly personnel and other persons who are not employees of the technical service department of PROTEC GmbH & Co. KG are requested to contact the local branch of PROTEC GmbH & Co. KG before assembly or service work is started.
  • Page 9: Device Description

    The instructions for use describe the performance characteristics and operation required for efficient and effective use of the PRS 500 F. Before working with the PRS 500 F, the complete instructions for use must be read, especially the Safety Instructions and the chapter Handling.
  • Page 10: Installation

    The installation of the PRS 500 F must be performed by PROTEC service department or a service company authorized by them. This X-ray system PRS 500 F must be installed in a shielded X-ray room that complies with the national regulations on radiation protection.
  • Page 11: Image Receptor Floor Stand

    For the diagnosis of inflammatory and expansive processes of the mucosa, cranial bones and paranasal extension. • For the diagnosis of the abdomen (e.g. acute abdomen, plain abdominal radiography, urethrogram, cystogram). PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 11 von 44...
  • Page 12: Indications And Contraindications

    In den Dorfwiesen 14 | 71720 Oberstenfeld Telephone: +49 (0) 7062 – 92 55 0 Fax: +49 (0) 7062 – 92 55 60 E-Mail: protec@protec-med.com Internet: www.protec-med.com PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 12 von 44...
  • Page 13: Safety Instructions

    Settings and calibrations that are not described in these instructions for use must be carried out using the technical description of the device by PROTEC service department or a service company authorized by them.
  • Page 14 In the case of pregnant women and children, it must be carefully considered whether an exposure is necessary. It should be avoided if possible. PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 14 von 44...
  • Page 15: General Safety Instructions

    Power Box must be switched off. 2.1.2 Device Operation In case of a malfunction, do not use the PRS 500 F anymore and notify PROTEC service department or a service company authorized by them. 2.1.2.1 Operating Type The PRS 500 F is not intended for continuous operation.
  • Page 16: Crushing And Collision Hazard

    It must be ensured that when operating the moving parts of PRS 500 F, no persons or objects are in the obvious danger area of the device. If not observed, it can result in personal injury or damage to the PRS 500 F or other objects.
  • Page 17 The user may need to take remedial measures such as relocating or realigning the device. The PRS 500 F is intended for use in an environment in professional health care facilities (e.g. clinics, surgery centres, physiology practices ...) PROTEC GmbH &...
  • Page 18: Control Elements And Displays

    Main Switch of the X-ray system The PRS 500 F is switched on and off via a mini console (e.g. of the Venus-32/50-R, see Fig. 3.1a), a normal control console (e.g. of the Venus-32/50-R, see Fig. 3.1b) of the X-ray generator or via the switch on the Power Box (see Fig.
  • Page 19: Emergency Stop Switches Of The X-Ray System

    Instructions for use Emergency Stop Switches of the X-ray system The PRS 500 F has the following emergency stop switches, which can be used to bring the device to an immediate standstill and disconnect it from the power supply. 3.2.1...
  • Page 20: Prognost Sh Touch

    If the operating unit is equipped with handles, the electromagnetic lock of one or more movements can be released by thumb pressure on the keys of the operating unit and the tube head can be brought into the desired position. PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 20 von 44...
  • Page 21: Foot Pedal

    There is an increased risk of injury if the X-ray tube carrier is not engaged! Detailed information please find in the accompanying User Manual PROGNOST SH. PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 21 von 44...
  • Page 22: Control Elements And Display Of Collimator

    Detailed information please find in the enclosed User Manual of the X-ray generator. Control elements of Bucky, Grid entity Detailed information please find in the enclosed User Manual. PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 22 von 44...
  • Page 23: Control Elements And Display Of Verticalx-Ray System Image Receptor Stand Provert

    Detailed information please find in the enclosed User Manual of the RAPIXX system. 3.11 Control elements and device displays of CONAXX 2 Detailed information please find in the enclosed User Manual of the CONAXX 2. PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 23 von 44...
  • Page 24: Handling / Operation

    At the X-ray generator operator console control panel, select the desire X-ray equipment (Bucky table with Bucky unit) by selecting the corresponding X-ray equipment/exposure technique switch. PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 24 von 44...
  • Page 25: Overtable Exposures

    Push the tube column to the lateral detector holder − Adjust the height of the tube arm or of the table that the light field is aligned to the panel PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 25 von 44...
  • Page 26: Operation At X-Ray System Image Receptor Stand Provert

    (Exposure every 30 seconds, otherwise tube may arc). See User Manual Generator and CONAXX 2 User Manual chapter 5.3. Detailed information please find in the enclosed User Manual of the X-ray tube. PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 26 von 44...
  • Page 27: Operation X-Ray Generator

    4.11.1 Switching the PRS 500 F on and off Switching on the PRS 500 F is switched on via the console of the generator. All system components are supplied with voltage via the generator. If the system contains a Power Box, the power is supplied via the Power Box.
  • Page 28: Safety And Maintenance

    (see also chapter 4.1). The PRS 500 F must no longer be used with patients if it shows signs of wear (e.g. metal abrasion, wear of insulations) or dangerous technical defects (e.g. torn cable, bent parts) or if the resulting image quality (e.g.
  • Page 29: Inspection And Maintenance

    (e.g. Dr. Schumacher Descosept Sensitive Wipes). All mechanical parts of the PRS 500 F, including accessories, may only be cleaned with a wipe disinfection with suitable surface disinfectants (e.g. Melsept® SF, 15 min. application time at 2% concentration).
  • Page 30: Product Service Life

    5.4.5 Product Service Life The PRS 500 F is designed for a service life of 10 years when used in accordance with the specifications and regular maintenance by PROTEC service department or a service company authorized by them. After the product has reached the end of its service life, further use is at your own risk.
  • Page 31: Power Supply

    Portable RF communication devices (radio devices) should not be used closer than 30 cm (12 in) to the marked parts and cables of the PRS 500 F. Failure to observe can lead to a reduction in the performance characteristics of the device.
  • Page 32: Guidelines And Manufacturer's Declaration - Electromagnetic Interference

    Guidelines and Manufacturer’s Declaration – Electromagnetic interference The PRS 500 F is intended for use in the electromagnetic environment specified below. The customer or the Operator of the radiographic system should assure that it is used in such an environment.
  • Page 33 1700 - 1990 DECT, 217 Hz 1970 LTE Band 1, 3, 4, 25, UMTS Bluetooth, Pulse modulation: 2450 2400 – 2570 WLAN, 802.11 217 Hz b/g/n, PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 33 von 44...
  • Page 34 PRS 500 F 5067-0-8002 Instructions for use RFID 2450, LTE Band 7 5240 Pulse modulation: 5500 5100 - 5800 WLAN, 802.11 a/n 217 Hz 5785 PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 34 von 44...
  • Page 35: Technical Data

    PRS 500 F 5067-0-8002 Instructions for use Technical Data Dimensions PATIENT PATIENT PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 35 von 44...
  • Page 36 PRS 500 F 5067-0-8002 Instructions for use PATIENT *OP - Anwender PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 36 von 44...
  • Page 37: X-Ray System Table Prognost F

    The Bucky, Grid entity and the measure chambers are connected to the generator. Detailed information please find in the enclosed User Manual, Technical Description of the Bucky, Grid entity. PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 37 von 44...
  • Page 38: X-Ray Tube Support Stand

    The Bucky, Grid entity and the measure chambers are connected to the generator. Detailed information please find in the enclosed User Manual, Technical Description of the PROVERT. PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 38 von 44...
  • Page 39: Attenuation Equivalent

    EN 60601-1-3 at 100kV and a first half-value layer thickness of 3.6 mm Al. 7.6.1 Protection Art and Protection Class The PRS 500 F corresponds to protection class 1 and contains applied parts type B (according to EN 60601-1). Automatic Cut-off Dose 7.7.1...
  • Page 40: Description Of Symbols, Labels And Abbreviations

    This symbol indicates the need to observe the instructions for use. This is provided in an electronic format (eIFU) on our website. www.protec-med.com/download Disposal instructions; WEEE (Waste of Electrical and Electronic Equipment) PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 40 von 44...
  • Page 41: Identification Label

    Horizontal movement of the X-ray system tube support, floor stand and vertical movement of the X-ray tube assembly Identification label Exemplary for PRS 500 F with X-ray generator 50 kW PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 41 von 44...
  • Page 42: Labels

    Labels on the X-ray system tube support, floor stand and columnX-ray system image receptor stand Product label Label on the front plate x-ray system table Company label PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 42 von 44...
  • Page 43: Position Symbols And Labels

    5067-0-8002 PRS 500 F Instructions for use Position symbols and labels Identification label Identification label PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 43 von 44...
  • Page 44: Abbreviations

    5067-0-8002 Instructions for use Abbreviations Millimetres Centimetres Pound Kilogram °C Degree -Celsius Hectopascal German Industry Standard European Standard CE-Mark Hertz Duty cycle Ampere Serial number PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 44 von 44...