PRS 500 E Basic diagnostic X-ray system Model/ID: 7069-9-8050L User Manual Ident. Nr. 5069-0-8002 PRS 500 E in analogue base configuration Version: 2.0 PROTEC GmbH & Co. KG Issued: 2019-08-27 In den Dorfwiesen 14, 71720 Oberstenfeld, Germany Subject to alterations...
Page 2
All sheets of this document contain proprietary and confidential information of PROTEC GmbH & Co. KG and is intended for exclusive use by current PROTEC GmbH & Co. KG customers. Copying, disclosure to other or other use is prohibited without the express written authorization of PROTEC´s law department.
Operation at the X-ray system table ........................22 4.1.1.1 Position of patients on the tabletop ....................... 22 4.1.1.2 Setting the X-ray unit on the mid moving Bucky, Grid entity ............22 PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 3 von 39...
Page 4
Environmental ....................................34 7.8.1 Environmental conditions during operation ....................34 7.8.2 Environmental Conditions for Shipping and Storage ................34 Description of symbols, labels and abbreviations ............35 PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 4 von 39...
Page 5
PRS 500 E 5069-0-8002 User Manual Symbols ....................................... 35 Identification label..................................37 Labels ........................................37 Position symbols and labels ..............................38 Abbreviations ....................................39 PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 5 von 39...
X-ray radiation. The intensity depends on the adjusted values of voltage, current and time. The radiation comes orthogonal out of the X-ray tube and is limited by the collimator. PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 7 von 39...
PROTEC GmbH & Co. KG Customers. Assemblers and other Customers not employed by nor directly affiliated with PROTEC GmbH & Co. KG technical services are directed to contact the local PROTEC GmbH & Co. KG office before attempting installation or service procedures.
This user manual describes the special features and operational aspects of the PRS 500 E, knowledge of which are required for efficient and effective use of the radiographic system. Prior to working with the PRS 500 E, it is required that the user reads the Safety Notes as well as the chapter regarding operation.
The radiographic system is primarily made of metal pieces. This has a main role in the weight of the device. The radiographic system PRS 500 E has a weight of 850kg (incl. Generator). Every technician is obliged to check the ground load. Also double bottoms and hollow floors have to be taken into account.
Intended user group The radiographic system PRS 500 E is exclusively designated for use by professional who are trained, in accordance with the corresponding national regulations, in the use of diagnostic X-Ry equipment and its proper intended use in connection with other medical devices, objects and accessories.
Adjustments and calibrations that are described within the user manual must be made, with the aid of The technical description for the system, by the PROTEC GmbH & Co. KG customer service department or a PROTEC GmbH & Co. KG authorized service technician.
Page 14
It’s not allowed to make any medical not indicated exposures on people. At pregnancy or children the question is if the exposure is really necessary. If possible it’s better to abandon it. PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 14 von 39...
Duty Cycle: S3 15% - maximum continuous operation of 1,5 minutes. 2.1.4 Operating personnel The radiographic system PRS 500 E with stand should only be operated by personnel who are trained in accordance with the corresponding national regulations in the use and operation of diagnostic radiographic systems.
2.1.6 Explosion protection These radiographic system PRS 500 E is not designated for use within areas with explosive hazards. 2.1.7 Radiation protection X-ray radiation can pose a hazard to patients and other people when the regulations regarding the operation of radiographic systems are not followed.
The user must take remedial measures such as implementation or reorientation of the device. The PRS 500 E is intended for the usage in a professional environment of the medical service (e.g. clinic, surgery centers, physiology offices …) PROTEC GmbH &...
Brake release movement of X-ray tube assembly around the horizontal support arm axis Brake release vertical movement of X-ray tube arm PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 18 von 39...
Brake release movement of X-ray tube assembly around the horizontal support arm axis Brake release vertical movement of X-ray tube arm Detailed information please find in the accompanying User Manual PROGNOST SH. PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 19 von 39...
Detailed information please find in the enclosed User Manual of the X-ray generator. Control elements of Bucky, Grid entity Detailed information please find in the enclosed User Manual. PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 20 von 39...
Detailed information please find in the enclosed User Manual of the RAPIXX system. Control elements and device displays of CONAXX 2 Detailed information please find in the enclosed User Manual of the CONAXX 2. PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 21 von 39...
Press the desired Anatomically Programmed Radiography (APR) switch or manually select the desired exposure factors. Start the exposure by pressing the switches for exposure preparation/exposure release. PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 22 von 39...
Detailed information please find in the enclosed User Manual of the PROGNOST E. Operation collimator Detailed information please find in the enclosed User Manual of the collimator. PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 23 von 39...
4.10 Function of the PRS 500 E 4.10.1 Switching On/Off the PRS 500 E Switching on the PRS 500 E happens via the control panel of the Generator. The Generator supplies every system component with power. On the Generator and the control panel will run a self-test when switching them on. After the self-test was successful the parameters will be displayed which can be saved under Organ-number #0.
If this happens properly the measured value will be displayed. If something is not running properly an Error message will be shown. Repeat this procedure for every measuring chamber. • The quickest exposure time of the Generator is 2ms. PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 25 von 39...
Cleaning The use of corrosive or abrasive cleaning agents as well as solvents is not allowed. These materials can cause damage to the outer surface of the radiographic system PRS 500 E or to the coating of the individual components.
5.3.5 Product life time The PRS 500 E has an expected product life of 10 years when used in accordance with the product specifications/ limitations and provided that maintenance through the PROTEC service department or a PROTEC authorized service provider has be completed. After reaching the life span the further usage of the device happens on own risk.
5.3.8 Disposal The X-ray system PRS 500 E contains different plastics, oils and heavy metal. At disposal of exchange parts or the whole system the current regulations have to be observed. Please contact the service or a company specialized for disposing the components.
CAUTION! Mobile HF-Communication devices shouldn’t be used closer than 30cm (12 Inch) to the marked parts and cables of the PRS 500 E. Disregarding this can cause a decrease in the performance features of the device. CAUTION! The X-ray generator integrated into the radiographic system PRS 500 E sends out electromagnetic waves during operation, which could cause interference with other devices.
The customer or the user of the radiographic system can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitter) and the radiographic system. PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 30 von 39...
PRS 500 E 5069-0-8002 User Manual Technical Data Dimensions PATIENT PATIENT PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 31 von 39...
PRS 500 E 5069-0-8002 User Manual PATIENT *OP – Operator X-ray system table PROGNOST E Table top dimension (L x B): 2260 mm x 755 mm, standard PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 32 von 39...
The Bucky, Grid entity and the measure chambers are connected to the generator. Detailed information please find in the enclosed User Manual, Technical Description of the PROVERT. Attenuation Equivalent WARNING! PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 33 von 39...
0,4 and < 0,5 Al mm according to EN 60601-1-3. Tested at 100 kV with a first half-value layer thickness (HVLT) of 3,7 mm Al. 7.6.1 Protection Art and Protection Class The PRS 500 E is consistent with a protection class 1 device and contains applicable parts Type B (according to EN 60601-1).Environmental conditions. Automatic cutoff dose 7.7.1 Analogue System The automatic cutoff dose is 2.5 µGy.
Homepage www.protec-med.com/download Notes on disposal; WEEE , Waste of Electrical and Electronic Equipment Protective ground (Earth) Caution: pinch-/crushing hazard for hands and fingers PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 35 von 39...
Page 36
Vertical movement of the X-ray tube arm Rotation of the X-ray tube assembly Horizontal movement of the X-ray system tube support, floor stand and vertical movement of the X-ray tube assembly PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 36 von 39...
Maximum allowable Patient weight (distributed load) for the tabletop (Composite-fibre tabletop). Maximum allowable Patient weight (distributed load) for the tabletop (Carbon tabletop). Labels on top of the tabletop and/or tube arm PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 37 von 39...
Product label Label on the front plate x-ray system table Company label Position symbols and labels At both ends of the rail Identification label PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 38 von 39 Identification label...
PRS 500 E 5069-0-8002 User Manual Abbreviations Millimetres Centimetres Pound Kilogram °C Degree -Celsius Hectopascal German Industry Standard European Standard CE-Mark Hertz Duty cycle Ampere Serial number PROTEC GmbH & Co. KG, In den Dorfwiesen 14, 71720 Oberstenfeld, Germany 39 von 39...