Datcon DT9102 B Operating Instructions Manual page 40

Intrinsically safe digital clock
Table of Contents

Advertisement

Sequence of operations 11. If you want to test synchronisation input, press ▲
Exit from the menu item 1. Press the MENU button, to exit from the Tests menu
Exit from the menu
40

button to change
(Its meaning: synchronisation input test)
12. Press ENTER button.
13. You will see on the display:
when the synchronisation input is in non-active state
(terminals 7 and 8 are, R > 200 kΩ)
or you will see on the display:
when the synchronisation input is in active state (terminals
7 and 8 are closed, R < 1 kΩ)
14. Press MENU button to exit this menu item:
you will see on the display the
mnemonic will be blinking.
15. Press the ▲ button to change the

the
mnemonic. (Its meaning: Temperature
measurement)
16. Press ENTER button.
17. You will see on the display the internal temperature of
the instrument in °C.
18. Press MENU button to exit this menu item: you will see
on the display: the
bee blinking.
item, and you see this:
(2. If you want to return to the tests, go on from point 3 of
the Sequence of operations.)
(3. If you don't want to exit the menu because you want to
perform further settings, you can select the desired menu
items by pressing the ▲ button.)
1. Press the MENU button. First the mnemonic
(Save), then the mnemonic
display. With this the storing of the settings is completed.
The instrument has exited the menu and goes on with the
measurement.
mnemonic to




mnemonic. The mnemonic will

.

(Exit) are shown on the
DT9102 B

mnemonic.
mnemonic. The

mnemonic to

20180905-V120180905