Interface Reconfiguration - Et system LAB-HP/E Series Manual

Table of Contents

Advertisement

Ext. Control: Computer
Example:
#1,ID
#22,GTR
#ALL,GTL
The status word can be read with the command
Bit
D15
Parity error
D14
Over run error
D13
Framing error
D12
Timeout error
D11
n/a
D10
n/a
D9
n/a
D8
n/a
D7
Parity enable
D6
Parity mode (1 = odd, 0 = even)
D5
Stop bit (1 = 2 Stop bits; 0 = 1 Stop bit)
D4
Data format (1 = 8 bit; 0 = 7 bit)
D3
n/a
 Table
D2
 Table
D1
 Table
D0
Table: Error messages
D2
D1
0
0
0
1
0
1
1
0
1
0
1
1
Interface parameters are configured via software using the command
<SS>
command
.

Interface Reconfiguration

In case, the user has forgotten the active setup, there are two ways of reconfiguring the interface:
sending the command
using the display to configure the interface  Interface Parameters
22
<STB>
or
Function
D0
Error
1
Syntax
0
Command
1
Range
0
Unit
1
Hardware
0
Read
<PCx>
from one of the other interfaces
<*STB?>
. The following functions are assigned to the bits:
<PCx>
. The settings can be saved with the
ET System electronic GmbH

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the LAB-HP/E Series and is the answer not in the manual?

Questions and answers

This manual is also suitable for:

Lab-hp/e 5kwLab-hp/e 10kwLab-hp/e 15kw

Table of Contents