Fluke 1630 Calibration Information Manual page 9

Earth ground clamp
Hide thumbs Also See for 1630:
Table of Contents

Advertisement

Step
Action
Function
13
14
15
16
17
Note
[1]
Disregard reading accuracy during these calibration steps.
The calibration adjustment is now complete.
Table 4. Calibration Adjustment Steps (cont.)
Meter
Source
Enter Cal Mode
~ A
Calibrator and 6-in.
~ A
wire loop
~ A
Calibrator and 3
turns copper
~ A
magnet wire, 6-in.
diameter
Applied
Adjust
Value
Harmonic displays
Press Honce when
1 A /
60 Hz
reading is stable.
Press Honce when
3 A /
60 Hz
reading is stable.
Press Honce when
10 A /
60 Hz
reading is stable.
Press Honce when
30 A /
60 Hz
reading is stable.
Earth Ground Clamp
Calibration Adjustment
Expected
Result/Instruction
[1]
[1]
[1]
[1]
9