Troubleshooting guide
Symptom
No/intermittent
electrical power

Footswitch LED

Console not
responding
to footswitch
and footswitch
connector indicator
is NOT illuminated
System fault LED
illuminated
Cause
Power cord not connected or
connected loosely at console
or electrical outlet
Note: The console will default
to power level 1 if power is
interrupted
Power not present at
electrical outlet
Console power switch not
illuminated
Footswitch not connected
properly
Footswitch obstructed
Footswitch inoperative
Power fault caused
by over-current
System over-pressure
condition out of tolerance
Remedy
Ensure that power cord is:
• Fully seated into the back of the console and
electrical outlet
• Not damaged and free of defects
• For replacement power cord contact Customer
Care at 1-800-876-1261 (USA only) or local
Smith & Nephew representative
• Ensure that the electrical outlet has power
• Connect to a different electrical outlet
• Press power switch; power switch should
become illuminated
• Observe red alignment dots on footswitch
connector and footswitch socket are properly
aligned
• Ensure the footswitch connector is fully inserted
• Ensure that there are no objects obstructing the
footswitch from being depressed or releasing
• Order replacement footswitch (66800472) from
Customer Care at 1-800-876-1261 (USA only) or
local Smith & Nephew representative
1.
Turn console OFF by pressing power switch
2.
Ensure power cord is connected to an
appropriate power source. If necessary try a
different electrical outlet
3.
Wait at least 5 seconds after turning OFF
4.
Turn console ON by pressing power switch
• Ensure the yellow high pressure tube on
handpiece is not kinked, obstructed or tangled
• If error light is still illuminated, replace
handpiece
• Contact Customer Care, 1-800-876-1261 (USA
only) or local Smith & Nephew representative to
arrange a return

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the VERSAJET II and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for smith&nephew VERSAJET II

  • Page 1 Troubleshooting guide Symptom Cause Remedy No/intermittent Power cord not connected or Ensure that power cord is: electrical power connected loosely at console • Fully seated into the back of the console and or electrical outlet electrical outlet Note: The console will default •...
  • Page 2 Symptom Cause Remedy Handpiece does not No/obstructed fluid supply • Ensure saline bag is full and fluid flows freely prime • Check that pinch clamp is fully open Note: Priming takes • Check high pressure tube for kinks, obstructions approximately 1 min or leaks at power level 10 •...

Table of Contents