Area velocity sensor

Advertisement

Quick Links

Area Velocity Sensor
Manual Body #69-4353-024
Copyright © 2021. All rights reserved, Teledyne ISCO
Revision E, August 2021
Use and Disclosure of Data Information contained herein is classified as EAR99 under the U.S. Export
Administration Regulations.
Export, reexport or diversion contrary to U.S. law is prohibited.
TIENet
Installation and Operation Guide
350
®

Advertisement

Table of Contents
loading

Summary of Contents for Teledyne TIENet 350

  • Page 1 Area Velocity Sensor Installation and Operation Guide Manual Body #69-4353-024 Copyright © 2021. All rights reserved, Teledyne ISCO Revision E, August 2021 Use and Disclosure of Data Information contained herein is classified as EAR99 under the U.S. Export Administration Regulations.
  • Page 3 Teledyne ISCO recommends that you read this manual completely before placing the equipment in service. Although Teledyne ISCO designs reliability into all equipment, there is always the possibility of a malfunction. This manual may help in diagnosing and repairing the malfunction.
  • Page 5 TIENet™ 350 Area Velocity Sensor Safety TIENet™ 350 Area Velocity Sensor Safety General Warnings Before installing, operating, or maintaining this equipment, it is imperative that all hazards and preventive measures are fully understood. While specific hazards may vary according to location and application, take heed of the following general warnings: WARNING...
  • Page 6 TIENet™ 350 Area Velocity Sensor Safety The equipment and this manual use symbols used to warn of Hazard Symbols hazards. The symbols are explained below. Hazard Symbols Warnings and Cautions The exclamation point within the triangle is a warning sign alerting you of important instructions in the instrument’s technical reference manual.
  • Page 7: Table Of Contents

    4.6 Contact Teledyne ISCO........
  • Page 8 TIENet® Model 350 Area Velocity Sensor Table of Contents Appendix A Replacement Parts List A.1 Replacement Parts Diagrams and Listings ......A-1 Appendix B Velocity Error Codes B.1 Introduction .
  • Page 9: Section 1 Introduction

    TIENet Model 350 Area Velocity Sensor ® Section 1 Introduction The Teledyne ISCO TIENet 350 Area Velocity Sensor measures flow stream average area velocity and liquid level. The Sig- nature Flow Meter uses this information to calculate the flow ®...
  • Page 10: 350 Velocity Operation

    TIENet® Model 350 Area Velocity Sensor Section 1 Introduction The 350 AV sensor is available with a 10m, 23m cable. For greater distances, external connection via conduit, and con- nection of additional TIENet devices, the TIENet Expansion Box is available. Bulk TIENet cable may also be used for greater dis- tances.
  • Page 11: Technical Specifications

    TIENet® Model 350 Area Velocity Sensor Section 1 Introduction 1.2 Technical Specifications Table 1-1 350 TIENet Sensor Specifications 1.9 3.3 15.2 cm 0.75 1.31 6.00 in Sensor Dimensions Standard Cable Length 10, or 23 m 16.4, 32.8, or 75.5 ft Cable Diameter 10.2mm ±0.254mm 0.402 in ±0.010 in...
  • Page 12: Optional Laserflow Applications

    TIENet 350 AV Sensor w/ 10m Cable ..................60-4354-113 TIENet 350 AV Sensor w/ 23m Cable ..................60-4354-114 TIENet 350 AV Sensor w/ 2ft Cable & TIENet Plug Connector..........60-4354-117 Cord grip fitting, " NPT, for TIENet cable ................209-0073-12 Bulk TIENet Cable, Cut to Length ..................
  • Page 13: Ordering Information

    When the system arrives, inspect the outside packing for any damage. Then carefully inspect the contents for damage. If there Instructions is damage, contact the delivery company and Teledyne ISCO (or its agent) immediately. WARNING If there is any evidence that any items may have been damaged in shipping, do not attempt to install the unit.
  • Page 14: Section 2 Installation

    TIENet Model 350 Area Velocity Sensor ® Section 2 Installation The Signature Flow Meter does not have to be mounted near the flow stream. You can install the flow meter itself at a convenient, protected location and route the sensor cable to the measurement point (maximum length of 305 meters or 1,000 feet).
  • Page 15: Reference Line Support

    TIENet® Model 350 Area Velocity Sensor Section 2 Installation 2.2 Reference Line The optional water-tight Signature Expansion Box enables a variety of Support configurations for adding length, as well as connecting multiple devices at once. The Expansion Box con- nects to a TIENet™ terminal strip in the Signature, and contains three additional strips inside, as well as a TIENet connection for an...
  • Page 16: Preparing The Signature Flow Meter

    Signature Flow Meter For Signature systems using the 330 Bubbler Module and/or the 2.3.1 External Desiccator TIENet 350 Area Velocity Sensor, the external desiccator dries the reference air for a pressure transducer and air supply for the bubbler. Signature bubbler systems will already have a desiccator installed.
  • Page 17: External Desiccator, Installed

    TIENet® Model 350 Area Velocity Sensor Section 2 Installation To air intake To reference port Figure 2-3 External desiccator, installed The desiccant cartridge requires periodic maintenance. Refer to Section 4.3 External Desiccator for instructions. Some Signature flow meters have a single piece of tubing Optimizing drying power installed between the reference port and the humidity connector, and a cap plug on the intake port.
  • Page 18: Connecting The Cable

    TIENet® Model 350 Area Velocity Sensor Section 2 Installation Intake port tubing Reference port tubing Humidity connector tubing Figure 2-4 Tubing configuration for optimal drying power 2.4 Connecting the Cable External TIENet devices such as the 350 are all electrically con- nected to the Signature flow meter in the same manner, usually using conduit or cord-grip cable fittings.
  • Page 19: Tienet Device Terminal Strips

    TIENet® Model 350 Area Velocity Sensor Section 2 Installation 1. Remove one of the 6-position plug-in terminal strip connec- tors from the case board. Figure 2-5 TIENet Device terminal strips 2. If using a cord-grip fitting, install the cable nut in the appropriate opening on the bottom of the Signature enclo- sure, securing it to the wall with the lock nut (concave side- facing wall).
  • Page 20: Tienet Device Terminal Connections

    TIENet® Model 350 Area Velocity Sensor Section 2 Installation Note The SHIELD wire is the bare drain emerging from the foil shi eld a roun d th e Y E LLO W an d B ROW N wire s. T he BRAID-DRAIN wire is the bare drain emerging from the sur- rounding braided shield inside the cable jacket.
  • Page 21 TIENet® Model 350 Area Velocity Sensor Section 2 Installation 5. Insert the reference tubing into the REF AIR port on the case board, pushing it down inside the silicon tubing. Becareful not to kink the reference tubing. Note: Remove the red cap from the reference port, on the case board next to the TIENet connector, being used by the 350 sensor...
  • Page 22: Installing The 350 Av Sensor

    TIENet® Model 350 Area Velocity Sensor Section 2 Installation 2.5 Installing the 350 AV Prior to mounting the sensor in the flow stream, check the dis- played level reading. In open air, the reading should be zero. If Sensor not, adjust the level to zero from the Level Adjustment screen (refer to “Setting the 350 Ex Level”...
  • Page 23: Mounting Rings

    For pipes up to 15" (38 cm) in diameter, stainless steel self-expanding mounting rings (Spring Rings) are available. For pipes larger than 15" in diameter, Teledyne ISCO offers the Scissors Rings (Universal Mounting Rings). Area velocity sensors can also be installed using primary measuring devices.
  • Page 24 TIENet® Model 350 Area Velocity Sensor Section 2 Installation Figure 2-11 Sensor Installed on a Spring Ring Attaching the Sensor to the Attach the 350 sensor to the ring either by using two 4-40x Ring countersink screws or by snapping the optional probe carrier to the ring.
  • Page 25: Scissors Mounting Ring

    For pipes larger than 15" in diameter, Teledyne ISCO offers the 2.5.4 Scissors Mounting adjustable Scissors Ring (also known as the Universal Mounting Ring Ring).
  • Page 26 Base Section Figure 2-12 Scissors Ring adjustment Note The hardware kit includes flat head screws and nuts.Teledyne ISCO strongly recommends bolting the assembled sections together before installation, using the holes provided for that purpose. This can greatly increase safety and prevent the assembly from being torn apart.
  • Page 27: Completing The Sensor Installation

    15 feet. For more information about the Street Level Installation System, contact your Teledyne ISCO representative. 2.7 Grounding Kit for Added protection from lightning damage is available with the grounding lug kit.
  • Page 28: Section 3 Setup And Programming For The Signature

    Signature 3.1 Configuring the To configure the Signature flow meter for operation with the System TIENet 350 sensor, press ) to access the top menu, MENU and select Hardware Setup. For all TIENet devices including the 350 Ex, select Smart Sensor Setup (TIENet).
  • Page 29: Character Grid

    TIENet® Model 350 Area Velocity Sensor Section 3 Setup and Programming for the Signature 3 5 0 Ve l o c i t y Done Cancel B C D E F G H I J K L M N c d e f g h i j k l m n o p ) - _ + = <...
  • Page 30 TIENet® Model 350 Area Velocity Sensor Section 3 Setup and Programming for the Signature Select Operation 1. Hardware Setup 2. Configure 3. Administration 4. Home Hardware Setup 1. Smart Sensor Setup (TIENet) 2. SDI-12 Setup 3. MODBUS Input Setup 4. MODBUS Output Setup 5.
  • Page 31 TIENet® Model 350 Area Velocity Sensor Section 3 Setup and Programming for the Signature 3.2 Measurement Setup From Measurement Setup (Figure 3-3), you can set up velocity measurement or make changes to the advanced settings. Note Refer to your Signature user manual for information about Flow Rate Input Setup and Volume Input Setup.
  • Page 32: Advanced Settings

    TIENet® Model 350 Area Velocity Sensor Section 3 Setup and Programming for the Signature The sensor is pre-programmed at the factory with the Advanced 3.2.2 Advanced Settings Settings for your application. Should your application require the addition of any correction factors, the Advanced button opens the Advanced settings window (Figure 3-4).
  • Page 33: Setting The 350 Level

    TIENet® Model 350 Area Velocity Sensor Section 3 Setup and Programming for the Signature The Level Adjustment screen is accessed via the Shortcuts menu 3.2.3 Setting the 350 Level on the Signature. From this screen, you can also update the display to show the current level of the stream.
  • Page 34: Section 4 Maintenance

    TIENet Model 350 Area Velocity Sensor ® Section 4 Maintenance 4.1 Maintenance The following tables are recommended maintenance checks to ensure proper operation. As site conditions may vary, increase the frequency of inspections as needed. Severity of humidity and likelihood of ragging (blockage at a spe- cific site) must be considered when determining regularly scheduled maintenance.
  • Page 35: External Desiccator

    Note Teledyne ISCO recommends checking the desiccant at least every 6 months, and changing/renewing the desiccant before the entire compartment has changed color.
  • Page 36: In Pipe Or Channel Sensor Cleaning

    TIENet® Model 350 Area Velocity Sensor Section 4 Maintenance If removal is difficult, screw the caps back in and unscrew again. Gently knock the caps and the cartridge against a hard surface to free any small particles in the threads, as these can hinder proper sealing and cause wear.
  • Page 37: Contact Teledyne Isco

    Direct or indirect contact with the outer diaphragm can permanently damage the 350 sensor. 4.6 Contact Teledyne If you have further questions about the installation, operation, and maintenance of your TIENet device, please contact our...
  • Page 38 TIENet Model 350 Area Velocity Sensor ® Appendix A Replacement Parts List A.1 Replacement Parts Replacement parts can be purchased by contacting Teledyne ISCO’sCustomerServiceDepartment. Teledyne ISCO Customer Service Department P.O. Box 82531 Lincoln, NE 68501 USA Phone: (800) 228-4373 or (402) 464-0231 FAX:(402) 465-3022 E-mail: isco.orders@teledyne.com...
  • Page 39 TIENet® Model 350 Area Velocity Sensor Appendix A Replacement Parts List REPLACEMENT PLUG SENSOR CAP KIT FOR 2100 SERIES AND TIENET DEVICES #60-9004-469...
  • Page 40: Appendix B Velocity Error Codes

    TIENet Model 350 Area Velocity Sensor ® Appendix B Velocity Error Codes B.1 Introduction Erroneous flow data can result from a number of factors. The area velocity system provides numbered error codes associated with the 350 Velocity data to assist in troubleshooting. The error codes are viewable using Isco Flowlink®...
  • Page 41: Signature Flow Data: Importing The .Ddp File

    TIENet® Model 350 Area Velocity Sensor Appendix B Velocity Error Codes A progress window will appear, displaying the filename, site name, device type, number of data types in the site file, and progress of the download. 5. When the two progress bars have completed, click Done to close the window.
  • Page 42: Viewing Velocity Error Codes In Flowlink

    TIENet® Model 350 Area Velocity Sensor Appendix B Velocity Error Codes B.3 Viewing Velocity Error In order to view error codes for velocity readings: Codes in Flowlink 1. In the Flowlink workspace, double-click the 350 Velocity data set. When the graph appears, click the Table View button.
  • Page 43 TIENet® Model 350 Area Velocity Sensor Appendix B Velocity Error Codes...
  • Page 44 Name and amount of Hazardous Substances or Elements in the product Hazardous Substances or Elements Component Name (Pb) (Hg) (Cd) (Cr(VI)) (PBB) (PBDE) Circuit Boards Display Wiring Internal Cables DC Motor Connectors Battery Solenoid valve Name and amount of Hazardous Substances or Elements in the product O: Represent the concentration of the hazardous substance in this component’s any homogeneous pieces is lower than the ST/ standard limitation.
  • Page 45 Warranty...

This manual is also suitable for:

Tienet 350 ex