TIENet® 310 Ex
Ultrasonic Level Sensor
Installation and Operation Guide
Ultrasonic Level Sensor
Manual Body #69-4313-010
Copyright © 2012. All rights reserved, Teledyne Isco
Revision E, March 2019

Advertisement

Table of Contents
loading

Summary of Contents for Teledyne TIENet 310 Ex

  • Page 1 TIENet® 310 Ex Ultrasonic Level Sensor Installation and Operation Guide Ultrasonic Level Sensor Manual Body #69-4313-010 Copyright © 2012. All rights reserved, Teledyne Isco Revision E, March 2019...
  • Page 3 Teledyne ISCO recommends that you read this manual completely before placing the equipment in service. Although Teledyne ISCO designs reliability into all equipment, there is always the possibility of a malfunction. This manual may help in diagnosing and repairing the malfunction.
  • Page 5 TIENet™ Model 310 Ex Ultrasonic Level Sensor Safety TIENet™ Model 310 Ex Ultrasonic Level Sensor Safety General Warnings Before installing, operating, or maintaining this equipment, it is imperative that all hazards and preventive measures are fully understood. While specific hazards may vary according to location and application, take heed of the following general warnings: WARNING...
  • Page 6 TIENet™ Model 310 Ex Ultrasonic Level Sensor Safety The equipment and this manual use symbols used to warn of Hazard Symbols hazards. The symbols are explained below. Hazard Symbols Warnings and Cautions The exclamation point within the triangle is a warning sign alerting you of important instructions in the instrument’s technical reference manual.
  • Page 7: Table Of Contents

    2.8 Troubleshooting TIENet 310 Ex USLS ....... . . 2-18...
  • Page 8 TIENet™ Model 310 Ex Ultrasonic Level Sensor Table of Contents List of Figures 1-1 Basic Signature monitoring system with 310 (mounting hardware not shown) ........1-1 1-2 310 Ultrasonic TIENet Sensor with unterminated leads (l) or TIENet plug (r) .
  • Page 9: Description

    TIENet® Model 310 Ex Ultrasonic Level Sensor Section 1 Introduction The Signature® Flow Meter uses the TIENet 310 Ex Device to provide non-contact liquid level measurement. The flow meter has built-in level-to-flow conversions that cover the majority of open channel flow measurement situations.
  • Page 10: 310 Ex Sensor Design

    TIENet® Model 310 Ex Ultrasonic Level Sensor Section 1 Introduction This non-contact measurement method reduces the frequency of maintenance, and is ideal for applications where the flow may contain chemicals, grease, silt, or suspended solids. 1.2 310 Ex Sensor Design The ultrasonic level sensor consists of a housing with a single transducer that is both pulse transmitter and echo receiver.
  • Page 11: Technical Specifications

    TIENet® Model 310 Ex Ultrasonic Level Sensor Section 1 Introduction 1.4 Technical Specifications Table 1-1 310 Ex TIENet Device Specifications 3.63"  x 4" tall (9.1cm  x 10.2cm tall) Sensor Dimensions Cable Length 10 or 23 meters standard (32.8 or 75.5ft) standard Mounting Attachment "...
  • Page 12: Accessories

    1.5.1 Ordering Information Options and accessories can be purchased by contacting Teledyne Isco’s Customer Service Department. Teledyne Isco Customer Service Dept.
  • Page 13: Tienet 310 Ex Ultrasonic Level Sensor

    " NPT, for TIENet cable ................209-0073-12 Barrier for Signature (SPA 2060) .................... 60-5324-060 Note Teledyne Isco uses FreeRTOS version 5.4.2 in its TIENet devices. In accordance with the FreeRTOS license, FreeRTOS source code is available on request. For more information, visit...
  • Page 14 TIENet® Model 310 Ex Ultrasonic Level Sensor Section 1 Introduction...
  • Page 15: Section 2 Installation And Setup For Signature

    TIENet® Model 310 Ex Ultrasonic Level Sensor Section 2 Installation and Setup for Signature The Signature Flow Meter does not have to be mounted near the flow stream. You can install the flow meter itself at a convenient, protected location and route the sensor cable to the measurement point.
  • Page 16: Waves

    TIENet® Model 310 Ex Ultrasonic Level Sensor Section 2 Installation and Setup for Signature 2.1.5 Waves Waves on the surface of the flow stream can deflect the ultrasonic signal, causing erroneous readings or total loss of signal. The Signature Flow Meter software is able to reject occasional readings that deviate substantially from normal.
  • Page 17: Tienet Device Terminal Strips

    TIENet® Model 310 Ex Ultrasonic Level Sensor Section 2 Installation and Setup for Signature 1. Remove one of the 6-position plug-in terminal strip connec- tors from the case board. Figure 2-1 TIENet Device terminal strips 2. If using a cord-grip fitting, install the cable nut in the appropriate opening on the bottom of the Signature enclo- sure, securing it to the wall with the lock nut (concave side facing wall).
  • Page 18: Tienet Device Terminal Connections

    TIENet® Model 310 Ex Ultrasonic Level Sensor Section 2 Installation and Setup for Signature Note The SHIELD wire is the bare drain emerging from the foil s h i e l d a r o u n d t h e Y E L L OW a n d B ROW N w i r e s. T h e BRAID-DRAIN wire is the bare drain emerging from the sur- rounding braided shield inside the cable jacket.
  • Page 19: Connecting To Signature Portable Via A Tienet Receptacle

    TIENet® Model 310 Ex Ultrasonic Level Sensor Section 2 Installation and Setup for Signature 6. Gently tug the cable downward, to remove any slack within the enclosure, taking care not to put any stress on the con- nection. 7. Tighten the cord grip sealing nut. CAUTION If you are using conduit instead of the cord-grip fitting, the con- duit must be sealed to prevent harmful gases and moisture...
  • Page 20: Sensor Installation

    TIENet® Model 310 Ex Ultrasonic Level Sensor Section 2 Installation and Setup for Signature • Area Velocity Sensor • 301 pH Interface • LaserFlow • 306 Sampler Interface Connecting a TIENet plug To connect the TIENet plug from the sensor to the TIENet Recep- to the Signature Portable tacle: 1.
  • Page 21: Dead Band

    The 310 Ex sensor can be mounted over the flow stream in various ways, depending on which method best fits the appli- cation. Optional equipment is available from Teledyne Isco for wall, floor, suspension, and horizontal mounting (see Section 1.5 Accessories). The sensor has a "...
  • Page 22: Installation In Hazardous Locations

    TIENet® Model 310 Ex Ultrasonic Level Sensor Section 2 Installation and Setup for Signature Wall Mount Approved for classified hazardous locations. Floor Stand Not approved for classified hazardous locations. Suspension (cable straightener recommended) Not approved for classified hazardous locations. Figure 2-7 Sensor mounting options 2.4 Installation in Read all labels carefully before installing the equipment! Hazardous Locations...
  • Page 23: Importantinformation Regarding "X" Marking

    Figures 2-9 and 2-10. Note There is TIENet barrier (a weatherproof box with terminals and power supply) available for hazardous locations (PN# 60-5324-060). Contact Teledyne Isco's SPA department for more details. Note Only the Wall Mount Bracket is approved for use in classified hazardous locations.
  • Page 24: Electrical Requirements

    (Li/Ri) of the 310 Ex device and integral cable. These parameters are established on the third party certification report and are available by contacting Teledyne Isco. 2.4.3 Ambient Environment Installation in designated hazardous areas must fall within the temperature range of -40 to +60°C, as specified on the serial tag...
  • Page 25: Sensor Labeling Regarding Hazloc Installations

    TIENet® Model 310 Ex Ultrasonic Level Sensor Section 2 Installation and Setup for Signature Temperature range Refer to Section 2.4.1 for important information about ’X’ marking. Ui = Maximum input voltage Ii = Maximum input current No objects of 12.5mm  or greater shall penetrate gap.
  • Page 26: Hazardous Location Installation Control Drawing-Atex

    TIENet® Model 310 Ex Ultrasonic Level Sensor Section 2 Installation and Setup for Signature Figure 2-9 Hazardous Location Installation Control Drawing-Atex 2-12...
  • Page 27: Hazardous Location Installation Control Drawing-Csa

    TIENet® Model 310 Ex Ultrasonic Level Sensor Section 2 Installation and Setup for Signature Figure 2-10 Hazardous Location Installation Control Drawing-CSA 2-13...
  • Page 28: Configuring The System

    Section 2 Installation and Setup for Signature 2.5 Configuring the To configure the Signature flow meter for operation with the System TIENet 310 Ex device, press ) to access the top MENU menu, and select Hardware Setup. For all TIENet devices including the 310 Ex, select Smart Sensor Setup (TIENet).
  • Page 29 TIENet® Model 310 Ex Ultrasonic Level Sensor Section 2 Installation and Setup for Signature Select Operation 1. Hardware Setup 2. Configure 3. Administration 4. Home Hardware Setup 1. Smart Sensor Setup (TIENet) 2. SDI-12 Setup 3. MODBUS Input Setup 4. MODBUS Output Setup 5.
  • Page 30: Measurement Setup

    TIENet® Model 310 Ex Ultrasonic Level Sensor Section 2 Installation and Setup for Signature 2.5.2 Measurement Setup From Measurement Setup (Figure 2-13 below), select Level Input Setup to define the measurement range. The Minimum Blanking Distance is the shortest distance from the sensor face (highest expected liquid level).
  • Page 31: Level Calibration

    TIENet® Model 310 Ex Ultrasonic Level Sensor Section 2 Installation and Setup for Signature 2.6 Level Calibration Although all other programming steps can be performed off-site, level must be set at the measurement site following installation, at ambient temperature. D – d = h (level) Once the sensor is installed over the flow stream, measure the present liquid level (see figure at left) and enter this value for Level, under Adjust Options.
  • Page 32: Firmware Updates

    Signature Flow Meter. Step-by-step instruc- tions for updating the firmware can be found in Section 2 of the Signature user manual. 2.8 Troubleshooting TIENet 310 Ex USLS Table 2-1 Troubleshooting: TIENet 310 Ex Ultrasonic Level Sensor Symptom Cause Action Not scanned...
  • Page 33: Replacement Parts

    Replacement parts are called out in the following illustrations. Refer to the call-out in the adjacent table to determine the part number for the item. Replacement parts can be purchased by contacting Teledyne Isco’s Customer Service Department. Teledyne Isco Customer Service Department P.O.
  • Page 34: Tienet 310 Ex Ultrasonic Level Sensor Replacement Parts

    TIENet® Model 310 Ex Ultrasonic Level Sensor A.1.1 TIENet 310 Ex Ultrasonic Level Sensor Replacement Parts Split mounting Nut ........................60-4313-007 Cap Sensor Connector ......................60-4314-011 310 Ex Ultrasonic sensor w/ 10m cable................... 60-4314-005 310 Ex Ultrasonic sensor w/ 23m cable................... 60-4314-006 310 Ex Ultrasonic sensor w/ Cut-to-length ................
  • Page 35 TIENet® Model 310 Ex Ultrasonic Level Sensor...
  • Page 36 TIENet® Model 310 Ex Ultrasonic Level Sensor...
  • Page 37 Compliance Statements Name and amount of Hazardous Substances or Elements in the product Hazardous Substances or Elements Component Name (Pb) (Hg) (Cd) (Cr(VI)) (PBB) (PBDE) Circuit Boards Display Wiring Internal Cables DC Motor Connectors Battery Solenoid valve Name and amount of Hazardous Substances or Elements in the product O: Represent the concentration of the hazardous substance in this component’s any homogeneous pieces is lower than the ST/ standard limitation.
  • Page 39 2014/30/EU - The EMC Directive 2014/35/EU - The Low Voltage Directive 2014/34/EU - The ATEX Directive 2011/65/EU – The RoHS Directive Manufacturer's Name: Teledyne Isco, Inc. Manufacturer's Address: 4700 Superior, Lincoln, Nebraska 68504 USA Mailing Address: P.O. Box 82531, Lincoln, NE 68501 Equipment Type/Environment:...
  • Page 41 The warrantor is Teledyne Isco, 4700 Superior, Lincoln, NE 68504, U.S.A. *This warranty applies to the USA and countries where Teledyne Isco does not have an authorized dealer. Customers in countries outside the USA, where Teledyne Isco has an authorized dealer, should contact their Teledyne Isco dealer for warranty service.