Agilent Technologies 5517B Operation And Service Manual page 62

Laser head
Hide thumbs Also See for 5517B:
Table of Contents

Advertisement

Figure 7- i. High Voltage Power Supply Schematic
Diagram
Paragraph 8-138:
Replace CAUTIONS with the following:
CAUTION
WHEN TAKING SIGNATURES, ENSURE THAT THE TEST-NORM SWITCH (A1S2) IS SET TO NORM AND
THAT HEATER TEST JUMPER IS IN THE OFF POSITION. DAMAGE TO THE HIGH VOLTAGE POWER
SUPPLY OR LASER TUBE ASSEMBLY COULD RESULT IF THESE INSTRUCTIONS ARE NOT ADHERED TO.
Add "Leave TEST-NORM A1S2 set to NORM." following "high voltage interlock." in step b.
Delete "• TEST-NORM = NORM"
Replace step u with the following:
u. If ending troubleshooting procedures at this point, disconnect power from HP 5517B Laser Head, set all test
jumpers to their left-most (normal) positions, return TEST-NORM switch A1S2 to NORM, and replace laser head
front panel and side covers.
Paragraph 8-140. Photodiode and Amplifier Troubleshooting
Add "Bypass high-voltage interlock by setting A1S2 to TEST" to the beginning of step a.
Change step f to read as follows:
f. If ending troubleshooting procedures at this point, disconnect power from HP 5517B Laser Head, set all test
jumpers to their left-most (normal) positions, return TEST-NORM switch A1S2 to NORM, and replace laser head
front panel and side covers.
Paragraph 8-142. Liquid Crystal Troubleshooting
Add "Set A1S2 to TEST" to the beginning of step a.
Replace step g with the following:
7-3

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

5517c5517d

Table of Contents