Health O Meter 2650KL User Instructions page 8

Professional digital wheelchair ramp scale
Hide thumbs Also See for 2650KL:
Table of Contents

Advertisement

2650KL ASSEMBLY INSTRUCTIONS (CONTINUED)
Assembling the 2650KL Scale (Refer to Figures 1-3)
Caution: It is recommended that the assembly be performed by two persons.
1. Place scale platform on a flat, level, and dry surface, and carefully turn the platform upside
down.
Caution: Ensure that the wheelchair ramps are held in place when moving the scale -
ramps can flip open.
2. Thread the (4) feet into the underside of the platform. Each foot should be completely hand
tightened into the platform. Back feet out 1-½ full turns. This will allow for proper load cell
operation and scale accuracy.
3. Carefully flip the scale over to its upright position on the floor.
Warning: Do not drop scale, as this can permanently damage the load cells.
4. Adjust the feet to level the scale. Be certain that the feet are not completely tightened into the
platform for proper load cell operation.
5. Open the scale to its operational position by folding open the wheelchair ramps.
6. Connect the load cell cable to the scale platform by inserting the load cell cable connector
into the load cell port (see Figure 2).
7. Ensure that the other load cell cable connector is securely plugged into the display assembly
load cell jack (located on the rear panel of the display assembly). The load cell jack is labeled
"LOAD CELL" (see Figure 3).
8. If the AC adapter will be used, connect the AC adapter cord plug into the AC power jack,
located on the back of the display assembly. The AC power jack is labeled "POWER" (see
Figure 3).
9. Connect the AC adapter to the power port, located on the back of the display assembly.
Caution: Do not plug the scale into a power source until the scale is fully assembled.
10. The scale is now assembled. Plug the AC adapter into a power source and / or insert the
batteries.
11. Carefully position the display assembly on the desk, table, or other flat surface.
Caution: Ensure that the load cell cable is adequately secured to prevent a tripping
hazard to the patient or user.
Note: Refer to "Display Assembly Wall Mounting Instructions", located in this User Instruction
Manual.
8

Advertisement

Table of Contents
loading

This manual is also suitable for:

2700kl

Table of Contents