Single Event Acquisition

To catch single events, you can perform a single shot acquisition (one-time screen update). To
set up the test tool for a single shot of the input A waveform, continue from step 3 again:
1. Use the cursor and E to select Single Shot.
The word MANUAL appears at the top of the screen indicating that the test tool is waiting for
a trigger. As soon as the test tool receives a trigger, the waveform is displayed and the
instrument is set to hold. This is indicated by the word HOLD at top of the screen. See
Figure
H
2. Press
The test tool stores all single shots in the replay memory. Use the Replay function to look
at all the stored single shots. See

N-Cycle Trigger

N-Cycle triggering enables you to create a stable picture of for example n-cycle burst waveforms.
Each next trigger is generated after the waveform has crossed the trigger level N times in the
direction that complies with the selected trigger slope.
To select N-Cycle triggering, continue from step 3 again:
1. Use the cursor and E to select On Trigger or Single Shot, jump to Trigger Filter.
sales@GlobalTestSupply.com
27.
to arm the test tool for a new single shot.
Figure 27. Making a Single Shot Measurement