Cursors On A Mathematical Result (+ - X) Waveform; Cursors On Spectrum Measurements; Rise Time Measurements - Fluke ScopeMeter 190 Series III User Manual

Hide thumbs Also See for ScopeMeter 190 Series III:
Table of Contents

Advertisement

Cursors on a Mathematical Result (+ - x) Waveform

Cursor measurements on, for example, an AxB waveform give a reading in Watts if input A
measures (milli)Volts and input B measures (milli)Amperes. For other cursor measurements on,
for example, a A+B, A-B or AxB waveform no reading will be available if the input A and input B
measurement unit are different.

Cursors on Spectrum Measurements

For a cursor measurent on a spectrum:
1. From the Spectrum measurement screen, press
g f
2. Use

Rise Time Measurements

For rise time measurements:
1. From Scope mode, press
2. Press 1 to highlight the Rise time icon.
3. For multiple waveforms, press 4 to select the required waveform A, B, C, D or M (if a
math function is active).
4. Press 3 to select MANUAL or AUTO (this automatically does steps 5 to 7).
h i
5. Use
A marker is shown at 90 %.
6. Press 2 to highlight the Fall time icon.
sales@GlobalTestSupply.com
to move the cursor and observe the readings at the top of the screen.
G
to move the upper cursor to 100 % of the waveform height.
Fluke-Direct
G
to show the cursor key label.
to show the cursor key labels.
®
ScopeMeter
Test Tool
Replay, Zoom and Cursors