10 MHz In, 10 MHz Out
The 10 MHz In BNC connector is used to synchronize the oscilloscope's
horizontal timebase system to a reference clock that you provide. The
clock that you provide must meet the following specifications:
• Amplitude: 178 mV peak to 1 V peak.
• Frequency: 10 MHz ±5 ppm high- quality sine wave or square wave.
To use an external reference clock, connect the external clock to the
10 MHz In BNC connector; then, in the Infiniium oscilloscope application's
Horizontal Setup dialog box (Setup > Horizontal), enable the 10 MHz External
Reference Clock.
You can use the 10 MHz Out BNC connector to send the oscilloscope's
10 MHz reference clock output signal to another instrument's reference
clock input.
Trig Out
Pulses corresponding to oscilloscope triggers can be sent to this BNC
output.
Aux Trig
You can set up the oscilloscope to trigger on the auxiliary trigger signal
connected to this BNC input.
Digital Channels Connector
MSO models include a connector for digital channel inputs.
Be sure to orient the cable as shown in the drawing on the back panel.
C AU T I O N
Inserting the cable incorrectly could bend the pins or cause errors.
Agilent Infiniium 90000 X-Series Oscilloscopes User's Guide
Using the Oscilloscope
29
2