RKI Instruments 35-3001A-07-03 Operator's Manual

Methane/hydrogen sulfide sample-draw detector

Advertisement

Quick Links

35-3001A-07-03
Methane/Hydrogen Sulfide
Sample-Draw Detector
Operator's Manual
Part Number: 71-0391
Revision: 0
Released: 8/8/17
www.rkiinstruments.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 35-3001A-07-03 and is the answer not in the manual?

Questions and answers

Summary of Contents for RKI Instruments 35-3001A-07-03

  • Page 1 35-3001A-07-03 Methane/Hydrogen Sulfide Sample-Draw Detector Operator’s Manual Part Number: 71-0391 Revision: 0 Released: 8/8/17 www.rkiinstruments.com...
  • Page 2 Typical calibration frequencies for most applications are between 3 and 6 months, but can be required more often or less often based on your usage. 35-3001A-07-03 Methane/Hydrogen Sulfide Sample-Draw Detector...
  • Page 3 Product Warranty RKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from defects in materials, workmanship, and performance for a period of one year from date of shipment from RKI Instruments, Inc. Any parts found defective within that period will be repaired or replaced, at our option, free of charge.
  • Page 4: Table Of Contents

    Parts List ..............25 35-3001A-07-03 Methane/Hydrogen Sulfide Sample-Draw Detector...
  • Page 5: Overview

    WARNING: When using the 35-3001A-07-03, you must follow the instructions and warnings in this manual to assure proper and safe operation of the 35-3001A-07-03 and to minimize the risk of personal injury. Be sure to maintain and periodically calibrate the 35-3001A-07-03 as described in this manual.
  • Page 6: Description

    AMP 1 AMP 2 Interconnect Terminal This End T o Strip Inlet F itting 3/4" Conduit Hub Particle F ilter INLET Inlet Fitting EXHAUST Exhaust Fitting Figure 1: Methane/Hydrogen Sulfide Sample-Draw Detector Component Location 2 • 35-3001A-07-03 Methane/Hydrogen Sulfide Sample-Draw Detector...
  • Page 7: External Components

    Pressure Switch Flowmeter Exhaust Hydrogen Flowmeter PCB Sulfide Restrictor Sensor Sensor Flow Control Valve Inlet Flow Baffle IR CH4 Sensor Pump Particle Filter Hydrophobic Filter Figure 2: Methane/Hydrogen Sulfide Sample-Draw Detector Flow Diagram 35-3001A-07-03 Methane/Hydrogen Sulfide Sample-Draw Detector • 3...
  • Page 8 A ball in the flowmeter column indicates the flow rate to the sensor. The flowmeter measures the flow in the range 0.2 to 2.0 SCFH (Standard Cubic Feet per Hour). The optimum flow rate is 1.0 SCFH. 4 • 35-3001A-07-03 Methane/Hydrogen Sulfide Sample-Draw Detector...
  • Page 9 4-wire cable mates to the IR CH sensor and retains it in the flow block. The cable is wired to the IR CH amplifier. The IR CH sensor PCB allows you to replace the sensor without disconnecting the wiring. 35-3001A-07-03 Methane/Hydrogen Sulfide Sample-Draw Detector • 5...
  • Page 10 The black test point in the upper left corner is the negative (-) test point and the red test point to the left of the zero and span pots is the positive (+) test point. 6 • 35-3001A-07-03 Methane/Hydrogen Sulfide Sample-Draw Detector...
  • Page 11 The black test point (CAL 2) in the upper left corner is the negative (-) test point and the red test point (CAL 1) in the upper right corner is the positive (+) test point. 35-3001A-07-03 Methane/Hydrogen Sulfide Sample-Draw Detector • 7...
  • Page 12: Installation

    6.50 Ø .30 x .50 (4X) M OUNTING 1.13 4.00 Pilot Fail Sample Drawing Gas Detector 8.88 8.50 35-3001 www.rkiinstruments.com 4.50 NOTE: Housing is 4.3 inches deep Figure 5: Outline and Mounting Dimensions 8 • 35-3001A-07-03 Methane/Hydrogen Sulfide Sample-Draw Detector...
  • Page 13: Connecting The Sample Lines To The Sample-Draw Detector

    1/4” O.D. rigid polypropylene, Teflon, or flexible polyurethane tubing may be used. RKI Instruments, Inc. recommends using either 1/4” O.D. x 1/8” I.D. or 1/4” O.D. x 0.170” I.D. tubing based on your length requirements. See “Specifications” on page 1 for maximum tubing lengths based on tubing size.
  • Page 14: Wiring The Sample-Draw Detector To A Controller

    Push 1/4” O.D. rigid polypropylene or rigid Teflon sample tubing into the fitting until it stops. Flexible polyurethane tubing may be used with an appropriate insert. RKI Instruments, Inc. recommends using either 1/4” O.D. x 1/8” I.D. or 1/4” O.D. x 0.170” I.D. tubing based on your length requirements.
  • Page 15 Connect the wires to the applicable detector/transmitter terminal strip at the controller as shown in Figure 6. Refer to the controller operator’s manual and the controller detector head specification sheet for the 35-3001A-07-03 for detector/terminal strip connections specific to the controller.
  • Page 16: Start Up

    If the display reading is 0 %vol for the methane channel and 0 ppm for the H S channel, start up is complete. The sample-draw detector is in normal operation. If the display reading is not 12 • 35-3001A-07-03 Methane/Hydrogen Sulfide Sample-Draw Detector...
  • Page 17: Maintenance

    “Calibration, IR CH Sensor” on page 21 or “Calibration, H S Sensor” on page 23, then continue this procedure. Set a voltmeter to measure in the millivolt (mV) range. 35-3001A-07-03 Methane/Hydrogen Sulfide Sample-Draw Detector • 13...
  • Page 18 13. Remove the voltmeter leads from the H S amplifier. 14. Close the housing door. 15. Store the calibration kit in a safe place. 16. Reconnect the sample tubing to the inlet fitting. 14 • 35-3001A-07-03 Methane/Hydrogen Sulfide Sample-Draw Detector...
  • Page 19: Troubleshooting

    If the fail condition continues, replace the sensor(s) as described in “Replacing Components of the Sample-Draw Detector” on page 16. If the fail condition continues, contact RKI Instruments, Inc. for further instruction. Slow or No Response/Difficult or Unable to Calibrate Symptoms •...
  • Page 20: Replacing Components Of The Sample-Draw Detector

    If you cannot set the correct flow rate, check the sample line for obstructions or kinks. If the calibration/response difficulties continue, replace the sensor as described later in this section. If the calibration/response difficulties continue, contact RKI Instruments, Inc. for further instruction. Replacing Components of the Sample-Draw Detector...
  • Page 21 There is a gasket at the bottom of the flow block. Be sure the gasket stays in place. Unplug the sensor from the H S amplifier. Verify that you are using the correct replacement sensor, then plug the sensor into the H 35-3001A-07-03 Methane/Hydrogen Sulfide Sample-Draw Detector • 17...
  • Page 22 Turn off power to the controller. Open the housing door of the sample-draw detector. Disconnect the filter from the rubber elbows on each end of the filter, then remove the filter from the sample-draw detector. 18 • 35-3001A-07-03 Methane/Hydrogen Sulfide Sample-Draw Detector...
  • Page 23: Adjusting The Low Flow Setting

    Decrease the flow very slowly by turning the sensor flow control valve’s knob counterclockwise and verify that the low flow alarm is 0.6 SCFH (±0.1). If the low flow alarm is set too high, turn the pressure switch adjustment screw slightly 35-3001A-07-03 Methane/Hydrogen Sulfide Sample-Draw Detector • 19...
  • Page 24: Removing The Particle Filter's Tubing Stub, If Necessary

    Screw the inlet fitting tube nut back onto the fitting body. See “Connecting the Sample Lines to the Sample-Draw Detector” on page 9 for instructions to install a new piece of tubing into the fitting. 20 • 35-3001A-07-03 Methane/Hydrogen Sulfide Sample-Draw Detector...
  • Page 25: Calibration Frequency

    Although there is no particular calibration frequency that is correct for all applications, a calibration frequency of every 3 months is adequate for most sample draw detector applications. Unless experience in a particular application dictates otherwise, RKI Instruments, Inc. recommends a calibration frequency of every 3 months for the H S detector and every 6 months for the infrared detector.
  • Page 26: Setting The Zero Reading

    Remove the voltmeter leads from the amplifier test points. Close the housing door. Follow the instructions in the controller’s operator’s manual to exit the calibration mode. Store the components of the calibration kit in a safe and convenient place. 22 • 35-3001A-07-03 Methane/Hydrogen Sulfide Sample-Draw Detector...
  • Page 27: Calibration, H 2 S Sensor

    5. If you verified a fresh air environment, proceed to the next section, Setting the Response Reading. Disconnect the sample tubing from the inlet fitting. Unscrew the regulator from the zero air calibration cylinder. 35-3001A-07-03 Methane/Hydrogen Sulfide Sample-Draw Detector • 23...
  • Page 28: Setting The Response Reading

    Remove the voltmeter leads from the amplifier test points. Close the housing door. Follow the instructions in the controller’s operator’s manual to exit the calibration mode. Store the components of the calibration kit in a safe and convenient place. 24 • 35-3001A-07-03 Methane/Hydrogen Sulfide Sample-Draw Detector...
  • Page 29: Parts List

    Regulator, demand flow, for 34 liter aluminum, 58 liter, and 103 liter calibration cylinders (cylinders with internal threads) 81-1055RK Regulator, demand flow, for 17 liter and 34 liter steel calibration cylinders (cylinders with external threads) ES-1537-H2S Hydrogen sulfide sensor 35-3001A-07-03 Methane/Hydrogen Sulfide Sample-Draw Detector • 25...