Page 2
Typical calibration frequencies for most applications are between 3 and 6 months, but can be required more often or less often based on your usage. 35-3001A-08-03 Methane/Carbon Monoxide Sample-Draw Detector...
Page 3
Product Warranty RKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from defects in materials, workmanship, and performance for a period of one year from date of shipment from RKI Instruments, Inc. Any parts found defective within that period will be repaired or replaced, at our option, free of charge.
WARNING: When using the 35-3001A-08-03, you must follow the instructions and warnings in this manual to assure proper and safe operation of the 35-3001A-08-03 and to minimize the risk of personal injury. Be sure to maintain and periodically calibrate the 35-3001A-08-03 as described in this manual.
Page 8
A ball in the flowmeter column indicates the flow rate to the sensor. The flowmeter measures the flow in the range 0.2 to 2.0 SCFH (Standard Cubic Feet per Hour). The optimum flow rate is 1.0 SCFH. 4 • 35-3001A-08-03 Methane/Carbon Monoxide Sample-Draw Detector...
Page 9
4-wire cable mates to the IR CH sensor and retains it in the flow block. The cable is wired to the IR CH amplifier. The IR CH sensor PCB allows you to replace the sensor without disconnecting the wiring. 35-3001A-08-03 Methane/Carbon Monoxide Sample-Draw Detector • 5...
Page 10
The black test point in the upper left corner is the negative (-) test point and the red test point to the left of the zero and span pots is the positive (+) test point. 6 • 35-3001A-08-03 Methane/Carbon Monoxide Sample-Draw Detector...
Page 11
The black test point (CAL 2) in the upper left corner is the negative (-) test point and the red test point (CAL 1) in the upper right corner is the positive (+) test point. 35-3001A-08-03 Methane/Carbon Monoxide Sample-Draw Detector • 7...
1/4” O.D. rigid polypropylene, Teflon, or flexible polyurethane tubing may be used. RKI Instruments, Inc. recommends using either 1/4” O.D. x 1/8” I.D. or 1/4” O.D. x 0.170” I.D. tubing based on your length requirements. See “Specifications” on page 1 for maximum tubing lengths based on tubing size.
Push 1/4” O.D. rigid polypropylene or rigid Teflon sample tubing into the fitting until it stops. Flexible polyurethane tubing may be used with an appropriate insert. RKI Instruments, Inc. recommends using either 1/4” O.D. x 1/8” I.D. or 1/4” O.D. x 0.170” I.D. tubing based on your length requirements.
Page 15
Connect the wires to the applicable detector/transmitter terminal strip at the controller as shown in Figure 6. Refer to the controller operator’s manual and the controller detector head specification sheet for the 35-3001A-08-03 for detector/terminal strip connections specific to the controller.
Enable alarms or place the controller in normal operation. CAUTION: Allow the sample-draw detector to warm up for 5 minutes before you continue with the next section, “Setting the Zero Reading.” 12 • 35-3001A-08-03 Methane/Carbon Monoxide Sample-Draw Detector...
Verify a display reading of 0 %vol for the methane channel and 0 ppm for the carbon monoxide channel at the controller. Investigate significant changes in the display reading. 35-3001A-08-03 Methane/Carbon Monoxide Sample-Draw Detector • 13...
Page 18
“Calibration, IR CH Sensor” on page 22. Unscrew the methane cylinder from the demand flow regulator. Remove the voltmeter leads from the IR CH amplifier. 14 • 35-3001A-08-03 Methane/Carbon Monoxide Sample-Draw Detector...
Probable causes • The sample-draw detector’s flow rate is too low because of an obstructed sample line, failed pump, etc. • The sample-draw detector is malfunctioning. • The sensor wiring is disconnected or misconnected. 35-3001A-08-03 Methane/Carbon Monoxide Sample-Draw Detector • 15...
If the fail condition continues, replace the sensor(s) as described in “Replacing Components of the Sample-Draw Detector” on page 16. If the fail condition continues, contact RKI Instruments, Inc. for further instruction. Slow or No Response/Difficult or Unable to Calibrate Symptoms •...
Page 21
Amplifier Controller Detector/Amp Terminal Terminal Strip Strip on Main PCB PWR/SIG “-” PWR/SIG “S” PWR/SIG “+” Table 3:Reconnecting the IR Methane Sensor to the Amplifier Amplifier Detector Detector Lead Terminal Strip DETECTOR “R” 35-3001A-08-03 Methane/Carbon Monoxide Sample-Draw Detector • 17...
Page 22
Unscrew the two screws in the upper right and lower left corners of the CO flow block then lift the CO amplifier and sensor off of the flow block. 18 • 35-3001A-08-03 Methane/Carbon Monoxide Sample-Draw Detector...
Page 23
Install the new filter. Be sure the side of the filter marked “INLET” is connected to the elbow that is connected to the inlet fitting. Turn on power to the controller. Turn on the controller. 35-3001A-08-03 Methane/Carbon Monoxide Sample-Draw Detector • 19...
If the low flow alarm is set too high, turn the pressure switch adjustment screw slightly clockwise. Repeat steps 3 and 4 if necessary. Use the sensor flow control valve to set the flow to 1.0 SCFH. Make sure the sample-draw detector’s Fail LED is off. 20 • 35-3001A-08-03 Methane/Carbon Monoxide Sample-Draw Detector...
Screw the inlet fitting tube nut back onto the fitting body. See “Connecting the Sample Lines to the Sample-Draw Detector” on page 9 for instructions to install a new piece of tubing into the fitting. 35-3001A-08-03 Methane/Carbon Monoxide Sample-Draw Detector • 21...
3 months is adequate for most sample draw detector applications. Unless experience in a particular application dictates otherwise, RKI Instruments, Inc. recommends a calibration frequency of every 3 months for the CO detector and every 6 months for the infrared detector.
Remove the voltmeter leads from the amplifier test points. Close the housing door. Follow the instructions in the controller’s operator’s manual to exit the calibration mode. Store the components of the calibration kit in a safe and convenient place. 35-3001A-08-03 Methane/Carbon Monoxide Sample-Draw Detector • 23...
5. If you verified a fresh air environment, proceed to the next section, Setting the Response Reading. Disconnect the sample tubing from the inlet fitting. Unscrew the regulator from the zero air calibration cylinder. 24 • 35-3001A-08-03 Methane/Carbon Monoxide Sample-Draw Detector...
Remove the voltmeter leads from the amplifier test points. Close the housing door. Follow the instructions in the controller’s operator’s manual to exit the calibration mode. Store the components of the calibration kit in a safe and convenient place. 35-3001A-08-03 Methane/Carbon Monoxide Sample-Draw Detector • 25...
Need help?
Do you have a question about the 35-3001A-08-03 and is the answer not in the manual?
Questions and answers