Configuration Mode - NCI 6712 User Manual

Point-of-sale interface scale
Table of Contents

Advertisement

Configuration Mode

Step-by-Step Instructions
for CONF Mode
Return to normal
operating mode by
pressing the SW-1
switch.
12
Model 6712 Point-of-Sale Interface Scale User's Manual
7. Press the TEST key to perform this test...
Indicates internal A/D raw count. When
weight is applied, the count value should
increase proportionately.
8. Press the TEST key to stop the test . . .
The remaining selections are for viewing
current settings only. You can scroll
through the menu to verify the settings,
but to make changes, you must enter
configuration or calibration.
The configuration (CONF) mode menu lets you
configure the scale to your specific application.
The items you can configure are as follows:
FILT (FILTER) – Adjusts the scale's response to
fit environment. Factory default setting is FAST.
BAUD (BAUD) – Allows selection of appropriate
baud rate and parity for computer interface.
Factory default setting is 96-N.
1. Press SW-1. . .
DIAG is displayed.
2. Press the ZERO key. . .
CONF is displayed.
3. Press the TEST key. . .
FILT is displayed.
4. Press the TEST key. . .
The present filter setting is displayed. the
two choices. When the proper selection